4 | Product description
44
D-R 290
Zero point measurement
Comparison standard
Measuring section
Reflector
Zero point
reflector
Transmitter
Turntable plate
50% splitter mirror
Receiver
Fig. 4.5: Zero point measurement
To check the correct operation of the D‑R 290 a checking cycle is performed at a peri
odic interval configurable from 0 to 720 hours (0 = no control cycle). During this cycli
cal check a zero point reflector is used to automatically check the contamination con
trol of the optical surfaces. The zero point and a reference are also measured auto
matically.
During the reference point measurement (
) with a
defined light attenuation is moved into the measurement beam. A check is then made
that the measured attenuation corresponds with the defined value. For zero point
measurement (
) the measurement beam is reflected directly in front of the
measurement medium. It thus passes through all the lenses and relevant optical com
ponents, just as in the real measurement procedure.
The measurement results for subsequent measurements are corrected to compensate
for the measured difference in value (contamination). If the contamination exceeds a
certain percentage (configurable between 0 and 10%), a warning is indicated. If the
contamination exceeds a higher percentage (>10%), an error message is indicated,
and the measured values from the device are thereafter no longer valid.
The evaluation electronics determines the transmission from the received measure
ment light and the intensity of the comparison light. From this the
, or
is measured. The extinction can be calibrated and
can be specified as dust concentration in mg/m³.
Furthermore for a known measuring path length the extinction coefficient and the
Koschmieder visual range can be calculated.
All the calculated values can be freely assigned to the four digitial measuring chan
nels.
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...