8 | Maintenance
D-R 290
161
Fig. 8.6: Example of a linearity test with scaling
Zero point check
The zero point check can be started and stopped by:
Setting path:
Device (D‑R 290 ) >
> Functions
> Maintenance /
Service Functions > Start zero point check
Device (D‑R 290 ) >
> Functions
> Maintenance /
Service Functions > Stop zero point check
During the zero point check the device measures the received intensity with the zero
point reflector swung into position, and compares it with the reference value. The out
put from the measuring channels is dependent in every case on the settings for the
channel (see section
Direct output of the opacity measured during the current measurement and during the
last zero point measurement can be found under:
Setting path:
Device (D‑R 290 ) >
> Specific parameters >
D‑R 290
Measured values > Zero point value [% OP
SP]
Contamination check
The D−R 290 automatically performs a contamination measurement once an hour.
The contamination measurement can also be performed manually via:
Setting path:
Device (D‑R 290 ) >
> Functions
> Maintenances /
Service Functions > Start
contamination check
Device (D‑R 290 ) >
> Functions
> Maintenance /
Service Functions > Stop contamination check
8.2.7
8.2.8
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...