7 | Checking / setting parameters with D‑ESI 100
D-R 290
133
Fig. 7.8: Common parameters D‑R 290 Variables setting II
Signal range
Each channel must be assigned a measurement range with a start and end
(D‑ESI 100 designation: signal range start / signal range end). The reference check
value must be specified based on the measurement range setting (e.g. 70% of the
measurement range) as must the range for the 4..20mA current output (4mA = signal
range start, 20mA = signal range end). The dimensions of the signal range settings
correspond to the settings in the type code (see section
Settings
It is specified for the channel whether during the check measurement the last meas
ured value should be retained or the reference check value output.
Zero range
Each channel has a zero range correction function. This allows specification that
measured values are set to <zero range (positive active) or measured values >zero
range (negative active) or measured values <>zero range equal to zero. This function
allows suppression of fluctuations around the zero point, negative values (positive ac
tive) or positive values (negative active).
1. "Zero range: positive active"
-> y = x if x > +1×|ZR|
y = 0 if x ≤ +1×|ZR|
(measured values ≤ +1×|ZR| are set to 0)
2. "Zero range: negative active"
-> y = x if x < -1×|ZR|
y = 0 if x ≥ -1×|ZR|
(measured values ≥ -1×|ZR| are set to 0)
3. "Zero range: positive active" and
"Zero range: negative active""
-> y = x if x > +1×|ZR| or x < -1×|ZR|
y = 0 if x ≤ +1×|ZR| and x ≥ -1×|ZR|
(combination of the two above functions)
7.5.3.5
7.5.3.6
7.5.3.7
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...