background image

 

 

 

VC110SDD 

User’s Manual

 

Page 12 of 16 

 
 

 

8. 

Refrigerator Loading 

 

IN ORDER TO KEEP INTERNAL TEMPERATURES OPTIMISED AND MINIMISE ENERGY 

CONSUMPTION, ONLY OPEN THE LID WHEN STRICTLY NECESSARY.

 

 

 

The VC110SDD refrigerator has a single compartment specifically for the storage of vaccines, see 

Section  4.  The  refrigerator  compartment  is  designed  to  maintain  a  stable  temperature  between 

+2

o

C and +8

o

C across an ambient temperature range from +5

o

C to +43

o

C .  There are no points 

within the vaccine compartment that drop below freezing and thus no danger of freezing vaccines.  

 

The  refrigerator  contains  stacking  baskets  designed  to  make  stock  management  easier  and 

encourage air circulation within the compartment.  

 

Do not remove these baskets and do not store vaccines outside of these baskets.

 

 

The wire baskets encourage the circulation of air by maintaining a small gap between the basket 

and the cabinet wall. Good air circulation will help the refrigerator to maintain an even temperature 

throughout the compartment. In order to save energy, try to keep the lid open for as short a time 

as possible.  

 

Stock management 

 

  Allocate a certain space for each different type of vaccine, so that they may be located rapidly. 

  Check the vaccine labels and adhere to any special requirements.  

  Always use before expiry date (store older vaccines on top).  

  Store the most frequently used vaccines in the top baskets for easy access.     

 

 

 

 

 

 

 

 

 

WARNING: ALWAYS FOLLOW THESE VACCINE STORAGE GUIDELINES 

 

Summary of Contents for VC110SDD

Page 1: ...DULAS Solar Direct Drive VC110SDD Refrigerator User s Manual Ref M VC110SDD ENG E ...

Page 2: ...YOUR DULAS SOLAR DIRECT DRIVE REFRIGERATOR 6 4 REFRIGERATOR COMPONENTS 7 5 MATERIAL SAFETY INFORMATION 8 6 END OF LIFE PRODUCT RECYCLING 9 7 REFRIGERATOR INSTALLATION 11 8 REFRIGERATOR LOADING 12 9 GREEN COMPRESSOR STATUS LED LIGHT 13 10 USER TRAINING 14 11 MAINTENANCE TASKS 14 12 VC110SDD SYSTEM LAYOUT 15 ...

Page 3: ...onnector Branch Plug 1 Solar Connector Branch Socket 1 Earthing Kit incl Earth Rod 25m Earth Cable Earth Clamp 1 MOUNTING SYSTEM see page 14 for references A to G A End Clamp 34 42mm Set 4 B Mid Clamp 28 42mm Set 2 C Front fixed leg 2 D C Rail 47 3 1530mm 2 E C Rail 47 3 750mm 2 F Rear adjustable leg 2 G Rail connector set 47 2 Coachscrews Rawbolts Washers 2 packs Screws 1 pack Wall Plugs 1 pack C...

Page 4: ...een given supervision or instruction Children must be supervised not to play with the appliance This refrigerator is for professional use only It is not designed for household use Refrigerator and power system are heavy please observe good manual handling procedures when moving the refrigerator and solar power system Only use suitable transport equipment Safe working Please observe safe working at...

Page 5: ...t expose it to rain WARNING This unit s cooling system contains flammable refrigerant Due to this fact the following information is particularly important Warning Do not damage the refrigeration circuit Ensure that sharp or pointed objects do not come into contact with the refrigeration circuit Warning Do not use mechanical devices or other means to accelerate the defrosting process other than tho...

Page 6: ...ng the day To maintain the temperature at night or when the sun is not shinning the refrigerator has a phase change lining This interior lining holds the exact temperatures required for safe vaccine storage and acts as a cold store storing energy that keeps the temperatures stable all day and night In order to store sufficient energy to keep vaccines safe for prolonged periods without any sunshine...

Page 7: ...VC110SDD User s Manual Page 7 of 16 4 Refrigerator Components ...

Page 8: ...de the walls of the refrigerator In normal use the user will never be aware of its presence This PCM is classified under regulation EC No 1272 2008 as a category 1 aspiration hazard and is classified under regulation 67 548 EWG 1999 45 EG as having the potential to cause physical harm R65 R66 At normal ambient temperatures this product will be unlikely to present an inhalation hazard because of it...

Page 9: ...efrigerator contain an organic phase change material that can be harmful and must be disposed of safely Please contact Dulas for advice Worn out devices must be professionally disposed of in accordance with local regulations and legislation Potentially hazardous components materials are Refrigerant gas R600a Phase change material PCM paraffins normal C5 20 Electrolytic capacitor Electronic compone...

Page 10: ...cycling facility Any device that you no longer require must be returned to our agent or find out about the approved collection and recycling facilities in your area Ignoring this European Directive may have potentially adverse effects on the environment and your health Solar modules contain materials that can be recovered and reused in either new PV modules or other new products Industrial recycli...

Page 11: ... the cable to allow access to the back of the refrigerator 4 Turn the PV isolator on by switching it to 1 Within 1 minute the compressor will start if there is sufficient sunlight solar irradiance of 125W m You will be able to hear the compressor operating and feel the condensers getting warm 5 Check that the refrigerator compartment temperature display on the front of the cabinet is operating If ...

Page 12: ...k management easier and encourage air circulation within the compartment Do not remove these baskets and do not store vaccines outside of these baskets The wire baskets encourage the circulation of air by maintaining a small gap between the basket and the cabinet wall Good air circulation will help the refrigerator to maintain an even temperature throughout the compartment In order to save energy ...

Page 13: ... solar power Normal operation The refrigerator warms up slightly overnight It is usual for the compressor to start running shortly after sunrise It will run continuously for a few hours to cool the refrigerator After this time the compressor will begin cycling on and off to maintain the correct internal temperature This will continue until the sun sets at the end of the day At night there is no po...

Page 14: ...only when required and lock after use Store the most frequently used vaccines in the top baskets for easy access Do not keep food or drink in the refrigerator Vaccine Storage Only store vaccines in the refrigerator Use the baskets provided Do not store vaccines outside the baskets Always store vaccine in its original packaging Always keep the same type of vaccine together and store your vaccine ne...

Page 15: ...VC110SDD User s Manual Page 15 of 16 12 VC110SDD System Layout ...

Page 16: ...DD User s Manual Page 16 of 16 For more information contact DULAS Ltd Dyfi Eco Park Machynlleth Powys SY20 8AX UK tel 44 0 1654 705055 fax 44 0 1654 703000 email solar dulas org uk website www dulassolar org ...