RFDU Operator’s Manual
16
RFDU OPERATING INSTRUCTIONS - Continued
Access Menu
The ACCESS menu allows a number of configuration
options to be “locked” or unable to be changed. When
selected the following screen will appear.
Each of the Recipe Config, Menu, Temp Mode
(C/F), Network and PHU Config can be locked or
unlocked.
For example, if “Recipe Config” is locked when the
Recipe Config option described previously is loaded
it will be in “VIEW MODE” only. Modifications such
as editing existing recipe settings, delete a recipe or
adding a new recipe will not be available.
Temp Offset
The Temperature Offset function will allow the
temperature offset to be adjusted per ZONE on the
selected RFDU.
After selecting the “TEMP OFFSET” function the
following screen will appear.
The screen displays the well or shelf configuration,
SHELF 1 and SHELF 2 actual bottom heatsink
temperatures and the current temperature offset
values. Each well or shelf can be individually
selected for editing.