background image

 

EXTRANEB MANOSONIC nebulizer 

 
 

1.

 

Choice of nasal plug 

Two nasal plugs are supplied to allow you to choose the size ensuring airtightness in nostrils. Introduce the soft tips 
in the nostrils, obstruct the back of the nasal plug. The pressure rise should be felt in the nose without leakage to 
the edge of the nostrils. If not, humidify the cones of the nasal-plug and try again. Otherwise, try a bigger nasal plug. 
 

2.

 

Instructions for use 

For  a  good  hygiene,  before  the  first  use,  sterilize  the  nebulizer 

except  nasal  plug

  in  boiling  water  (20  minutes). 

Connect the tube to the "Pressure" outlet of the device and to the inlet (1) of the nebulizer. Connect the big tube to 
the "Vibrations" outlet of the compressor and to the inlet (7). Fix the chosen nasal plug to the aerosol outlet (6). 
Open the upper part of the body of the jet nebulizer unscrewing it by ¼ revolution. 

Check that the jet-pipe (4) is 

correctly  inserted  on  the  bottom  part  (3)

.  Pour  the  solution  in  the  tank  (2)  without  overfilling  the  mark  "FULL"  

(12 ml, indicative value). Place back the upper part of the body of the nebulizer screwing it by ¼ revolution. Put the 
device  ON.  A  light  mist  of  aerosol  must  come  out  of  the  nebulizer.  If  not,  check  that  the  jets  (3)  or  (4)  are  not 
clogged and that the upper jet (4) is well positioned down on the bottom part (3). 
 
 

See instructions for use of the device. 

 

 

3.

 

Maintenance and cleaning 

After each inhalation session, clean the jets (3) and (4) by using the nebulizer with water. Then, 

unplug the tubes 

coming from the compressor

. Wash the nebulizer with hot water and let it dry. Do not use any metal wire to clean 

the jets (3) and (4). 
 

4.

 

Sterilization 

Dismantle the body of the nebulizer according to the picture.  
Nebulizer and tubes can be sterilized in boiling water (20 min).  

Nasal plugs can only be cold sterilized

 

5.

 

European Standard EN 13544-1 (with AMSA compressor) 

Maximum volume filling : 12 mL 
Minimum volume filling : 3 mL 
Performances for a 4 mLvolume: 
Aerosol output (a): 

0.93 mL NaF 1% 

Aerosol output flow (b):  0.09 mL/mn 
MMAD (c): 

4.5 µm 

(a)

 

Solution volume inhaled by the patient during one nebulization session. 

(b)

 

Solution volume inhaled by the patient during 1 minute of nebulization. 

(c)

 

Mass Median Aerodynamic Diameter (size of the aerosols particle). 

See the particles size rate curve in the device instructions for use. 
 

For hygienic purpose, a brand new EXTRANEB MANOSONIC nebulizer must be given to each patient.  
This nebulizer can be used with a MANOSONIC compressor of DTF MEDICAL.  
Expected shelf life of accessory: 1 year. 

 
 
 

 

GB 

Reviews: