unit into the charge device, the yellow indicator will enlightened, when 

charge finish, the green indicator will enlightened and the yellow one 

will off. The green light is the power indicator of the charging finish. If a 

battery error is detected, the indicator of the charge device will wink.

6.7 When operating finish, please clean the optical fiber with calico to 

avoid infecting the light intensity.

6.8 This equipment will turn off automatically if don't any action within 

2 minutes, turn it on by press any button.

6.9 The effective light intensity of this equipment is much more higher 

than Halogen Lamp, The solidified depth of the curing light composites 

resin for 10 seconds will not less than 4mm.

6.10 This equipment has the function of light intensity measuring. 

Connect the Adapter of the light intensity measure device correctly, 

aim the top of optical fiber at the sensor, turn on the main unit to emit 

blue light, the green indicator of the light intensity measure device will 

indicate the value of the light intensity.

6.11 The optical fiber should be sterilized for 4 minutes with 134

 and 

2.0bar~2.3bar (0.20MPa~0.23MPa) before each use.

6.12 The curing light is equipped with over-heat protection system. It can 

continuously work 200s, For example, continuously operate the curing 

light for 10 times under 20s working mode (even the curing light works 

less than 20s, it is counted as a full operation), then it will come into 

over-heat protection status. And only after 2-minute sleep, it can restart 

working 200s continuously.

7. Light intensity measurement 

7.1 Connect the output plug of power adapter to the plug of DC5.0V in 

the pedestal.

7.2 Choose general mode and aim the optical fiber at the measurement 

point, press on / off button, the present light intensity is displayed on the 

indicator of pedestal.

Summary of Contents for Lux E

Page 1: ......

Page 2: ...nting 3 6 Operation 4 7 Light intensity measurement 5 8 Precaution 6 9 Contraindication 6 10 Maintenance 6 11 Troubleshooting 7 12 Packing list 8 13 Transportation 8 14 Storage and transportation 8 15...

Page 3: ...and ultrasurgery etc 2 Principle and Application 2 1 LUX E adopts the principle of ray radiation to solidify the light sensitive resin by shooting at it in a short time 2 2 This product is used for de...

Page 4: ...2...

Page 5: ...here pressure 70kPa to106kPa 4 5 Size 40 50 260mm 4 6 Net weight 147g 4 7 Power consume 8W 4 8 Protection type against electrical shock Class 4 9 Protection degree against electrical shock Type B 4 10...

Page 6: ...the mode button to set the working mode the corresponding indicator will on when a mode set 6 1 1 Full power the blue light radiates in full power 6 1 2 Ramping the blue light power increases stronger...

Page 7: ...ntensity measure device correctly aim the top of optical fiber at the sensor turn on the main unit to emit blue light the green indicator of the light intensity measure device will indicate the value...

Page 8: ...y in the condition of cool and ventilated Please make sure of pressing out the buckle between the main unit and the pedestal otherwise the battery charging might be failed because of the poor contact...

Page 9: ...ttery is out of power 2 Faulty of battery 3 Battery is protected 1 C h a n g e a n e w battery 2 Change a new battery 3 Charge The screen shows Er Faulty of main unit Repair The screen shows E1 Low ba...

Page 10: ...list The components of the machine are listed in the packing list 13 Transportation 13 1 Excessive impact and shake should be prevented in transportation Lay it carefully and lightly and don t invert...

Page 11: ...during transportation 14 6 Keep it away form the sun rain or snow during transportation 15 After service From the date this equipment has been sold base on the warranty card we will repair this equip...

Page 12: ...ocuments Authorised Representative in the EUROPEAN COMMUNITY 18 Environmental protection Please dispose according to the local laws 19 EMC Declaration of conformity The device has been tested and homo...

Page 13: ...11 LUX E LUX E LUX E LUX E LUX E LUX E...

Page 14: ...12...

Page 15: ...hout further notice The pictures are only for reference The final interpretation rights belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO LTD The industrial design inner structure etc have claimed for...

Page 16: ...ZMN WI 09 132 V1 0 20181120...

Reviews: