DSO 15001K User Manual Download Page 4

How to Use

Display and Controls

Connections

Power Supply: 

Connect 9V DC power supply to the 5.5x2.1mm

jack at bottom (center positive). Power supply
voltage must be in the range of 8 - 10V.

Probe: 

Connect probe to the BNC connector at top.

Power supply voltage must not exceed

10V. Otherwise it may damage the ICs

inside.

Attention

1.

Allowed maximum signal input voltage

is 50Vpk (100Vpp) with the clip probe.

2.

Operations

[V/DIV]: 

Select sensitivity or vertical position. The selected parameter indicator will be highlighted.

[SEC/DIV]: 

Adjust the parameter seleted (highlighted). Short press toggles 

 mode.

Fast Adjustment

[TRIGGER]: 

Enter HOLD state (freeze waveform). Press it again will de-freeze.

Couple switch: 

Set couple to DC, AC, or GND. When GND is selected the scope input is isolated from input 

signal and connected to ground (0V input). 

Analog bandwidth
Sensitivity range

Resolution
Record length

Max realtime sample rate

Timebase range

Max input voltage
Input impedance

Power supply
Current consumption
Dimension
Weight

1MSa/s
0 -- 200KHz
5mV/div - 20V/div
50Vpk (1X probe)
1M ohm/20pF
12 bits
1024 points 

500s/Div -- 10us/Div

9V DC (8 

 10V)

~120mA @ 9V 
105 x 75 x 22mm
100 gram (without probe and PS)

Trigger modes
Trigger position

Auto, Normal, and Single

Center of buffer

Specifications

Save Waveform

Press [ADJ] & [

] buttons simultaneously. The currently displayed waveform

will be saved to EEPROM. The existing data in EEPROM will be over-written.

SEC/DIV

www.jyetech.com

JYE Tech Ltd.

Tel. +86-0773-2113856

Tech forum: www.jyetech.com/forum

Page 4

Basic Button Functions

 

[OK]: 
[ADJ]: 

Select timebase or horizontal position. The selected parameter

 will be highlighted.

 indicator

Select trigger mode, trigger level, and trigger edge. The selected parameter

 will be highlighted.

 indicator

More Functions

 

VPos Alignment

Measurements
ON/OFF

Functions

Operations

Set Couple Switch to GND position. Hold down [V/DIV] button for about 3 seconds.

Couple

Switch

Test Signal

output

BNC

Connector

Connectors

for Power Supply
(5.5 x 2.1mm)

Power 

Switch

HOLD RUN

Button

/

Oscilloscope

State

Vertical

Position
Indicator

Sensitivity

(V/div)

Couple

Timebase

(s/div)

Trigger

Mode

Trigger

Slope

Trigger

Level
Indicator

Trigger Level

Readout

Horizontal

Position
Indicator

Sensitivity

Button

Timebase

Button

Trigger

Button

Sen sitivity

Bu tton

Adjustment

Dial

(Signal Input)

Default Restore

Recall Waveform

Center HPos

Center Trigger

Level

Hold down [OK] button for about 3 seconds. This will turn ON or OFF on-screen display of 

measurements including Vmax, Vmin, Vavr, Vpp, Vrms, Freq., Cycle, Pulse width, and Duty cycle.

Press [

] & [

] buttons simultaneously. 

ADJ

Trigger

Recalled waveform is always displayed

in Hold state.

Hold down [SEC/DIV] and [TRIGGER] buttons simultaneously for about 3 seconds. 

Hold down [SEC/DIV] button for about 3 seconds. This will make the data at the center
of capture buffer displayed.

Hold down [TRIGGER] button for about 3 seconds. This will set the trigger level t0 the
medium value of signal amplitude.

Bad V+
Bad V-

Problems

Possible Causes

Connector J7 defective.

V1 does not close

to 0V

Bad AV+

Troubleshooting

1

2

Diode D2 open or damaged.

1

Bad C12 and/or C13.

2

U5 (7660) bad soldering or defective.

Bad AV-

1

R27 bad soldering or wrong value.

Hint: Check with R27 disconnected would let you know the issue is caused by load or source.

2

Shorts between AV- and ground.

1

R26 bad soldering or wrong value.

2

Shorts between AV+ and ground.

1

SW1 not set to GND position.

2

Bad soldering on R1 and/or R2.

3

Bad soldering on U1.

V2 does not close

to 0V

1

SW1 not set to GND position.

2

Bad soldering on R3 and/or R4.

3

Bad soldering on U1.

V3 does not close

to 0V

1

Bad soldering on U1 and/or U2.

2

Bad soldering on R5 and/or R6.

Bad V4

1

Bad soldering on R13, R14,  and R15.

No Trace

1

Incorrect V4. If V4 is correct perform factory default restore as described in 

2

Make sure trigger mode is AUTO and timebase is 1ms. Hold down [SEC/DIV] and [TRIGGER]

buttons simultaneously for 3 seconds.

2

www.jyetech.com -

JYE Tech Ltd.

(This becomes 

trigger state for
firmware version

-055 or later)

About Trigger State

The trigger can have three states including Holdoff, Waiting, and Trigged. They are explained below.

Holdoff:

Waiting:

Trigged:

Trigger is disabled until a portion of  sample buffer prior to a trigger point is filled with raw data.
Trigger is waiting for a valid signal slope.
A valid signal slope has been detected and registered.

Rolling Mode

When timebase is set to 50ms or slower and trigger mode is set to AUTO the scope will automatically switch to 

 where waveform shifts from right to left constantly. The trigger is disabled under this mode.

Rolling Mode

Fast Adjustment

Short press of [ADJ] toggles 

 mode on and off for VPos, HPos, and

Trigger Level. A 

>>

 sign appearing at top of screen indicates 

 is ON.

Fast Adjustment

Fast Adjustment

JYETech 

WWW.JYETECH.COM