- XI -

2) CALIBRATION METHOD

IF “SET/CAL” IS PRESSED ON S&C MODE, THE INDICATOR SHOWS “D XX”.
THE CALIBRATION IS DIVIDED TO 7 STEPS, AND THEN WHEN EACH STEP
IS DONE IN ORDER, AND THE LAST “END” IS INDICATED, THE PERFECT
CALIBRATION GETS TO BE FINISHED.
BE SURE TO USE “SET/CAL” KEY FOR THE PROGRESS OF EACH STEP,
AND “CLR” KEY FOR MOVING MODE TO THE FORMER STEP.

1. STEP : SETTINT-UP THE VALUE OF 1 SCALE (D)

2. STEP : SETTING UP THE WHOLE WEIGHT (CAPA)

USE THE NUMERICAL KEY AND PERFORM THE INPUT.
IF “SET/CAL” KEY IS PRESSED IN THE COMPLETION OF INPUT, IT IS
ENTERED INTOMEMORY, AND AT THE SAME TIME, IS TOWARD THE NEXT STEP.

3. STEP : IT INDICATES DEAD LOAD. (DEAD)

THE INDICATOR SHOWS THE CURRENT ZERO STATE, AND THEN DEVICE SO AS
THAT THIS VALUE CA HAVE THAT BETWEEN 1000-9000. (THE VALUE OF ZERO
DESIRALE IS 5000.)

4. STEP : ENTER THE INPUT OF WEIGHT OF SCALE FOR SPAN ADJUSTING (SPAN).

HOWEVER, BE SURE TO SET UP THE VALUE WHICH IS SIMILAR TO TOTAL.
WEIGHT OR IS MORE THAN 80% OF TOTAL WEIGHT AT LEAST.

5. STEP : SHOWING “UP” IN THE INDICATOR. (UP)

PUT THE SPAN ADJUSTING SCALE ON LOAD CELL.

6. STEP : SPAN ADJUSTING CONSTANT CALCULATION

IT CALCULATION THE CONSTANT AND SHOWS THE CONSTANT VALUE ON THE
INDICATOR. IF THIS PRICE IS LARGER THE “2.5, XXXX”, THE INDICATOR
SHOWS ERR-6.
THE CONSTANT VALUE DESIRABLE IS THAT BETWEEN 0.5-1.
IN CASE THE CONSTANT FOR SPAN ADJUSTING IS DELELTED, THE CALIBRATION
IS NOT DONE, AND THEN THE WEIGHT CAN BE ADJUSTED AS IT MEMORIZES
THE CONSTANT FOR SPAN ADJUSTING. (TO ADJUST THE WEIGHT IS F-99)

7. STEP : THE CALCULATION OF CONSTANT IS COMPLETED, AND THEN THE CONSTANT

VALUE IS INDICATED. AND THE, “END” MSG IS INDICATED, AMD IF THE SCALE
IS TAKEN OFF, AND “SET/CAL” KEY IS PRESSED, IT IS OPERATED TO
NORMAL MODE (USER MODE).
DIP SW7-8 IS ADJUSTED FOR LOAD CELL GAIN ADJUSTING.

Summary of Contents for 1260

Page 1: ... I Weighing Indicator Ds 1260 Operation Manual 혼을 담은 제품으로 귀사의 만족스런 파트너가 되겠습니다 감사합니다 DS LOAD CELL ...

Page 2: ... II ...

Page 3: ... IT IS LIGHTED SP 1 IN CASE WEIGHT VALUE IS LARGER THAN SP1 IT BECOMES LAMP ON SP 2 IN CASE WEIGHT VALUE IS LARGER THAN SP2 IT BECOMES LAMP ON SP 3 IN CASE WEIGHT VALUE IS LARGER THAN SP3 IT BECOMES LAMP ON SP 4 IN CASE WEIGHT VALUE IS LARGER THAN SP4 IT BECOMES LAMP ON 2 DISPLAY KEY BOARD WEIGHT DISPLAY IT INDICATES THE WEIGHT KEY BOARD IT IS USED WHEN SP1 SP2 SP3 AND SP4 ARE SET ...

Page 4: ...E ADJUSTED BEFORE THE POWER IS APPLIED BE SURE TO CONFIRM THE USABLE VOLTAGE AND THEN WHEN THE CHANGE OVER IS REQUIRED OPEN THE CASE AND THEN ADJUST THE CHANGE OVER SWITCH NEXT TO THE FOLLOWING TRANSFORMER NOTE IT IS ADJUSTED TO 110V IN THE DELIVERY OF PRODUCTS 6 RELAY OYT OUTPUT SP1 SP2 SP3 SP4 7 IN PUT INPUT GROSS NET TARE HOLD ZERO 8 RS232C CURRENT LOOP PORT RS232C CURRENT LOOP TO BE OPTION 9 L...

Page 5: ...ITH POWER ON MODE THE LETTERS OF TEST IS INDICATED IN THE DISPLAY IN THIS TIME WHEN 3 IS PRESSED AGAIN THE INDICATOR SHOWS SETUP CALIBRATION S C MODE 2 IN CASE IT WILL BE ENTERED FROM NORMAL MODE TO S C MODE WHEN SET CAL IS CONTINUOUSLY PRESSED MORE THAN 3 SECONDS IT BOCOMES S C MODE AND THE INDICATOR SHOWS THE LETTERS LIKE THE ABOVE 1 SECTION THIS MEANS THAT IT ASKS USER TO SELECT WHICH MODE OF S...

Page 6: ...ZERO SCOPE F06 DIGITAL FILTER 0 9 MOTION REDUCTING F07 VESSEL KEY OPERATION MODE STABLE NON STABLE OPERATION MODE F08 ZERO KEY OPERATION MODE STABLE NON STABLE OPERATION MODE F16 HOLD MODE PEAK SAMPLE MODE F17 HOLD INPUT MODE PULSE LEVEL F18 SAMPLE HOLD MODE MANUAL AUTOMATIC F20 PARITY MODE EVEN ODD NO PARITY BIT F21 TRANSMISSION MODE STREAM AUTO PRINT PRINTKEY F22 STABLE NON STABLE MODE STABLE NO...

Page 7: ...NUMBER SETTING UP METHOD THE NUMBER OF FUNCTION DESIRABLE IS INPUT WITH THE NUMERICAL KEY AND THEN CLR KEY IS PRESSED THE CURRENTLY SET UP VALUE IS INDICATED IN THE LAST 2 FIGURE AND THE NUMBER WHICH INTENDS TO BE CHANGED IS INPUT AND THEN SET CAL KEY IS PRESSED IF SO IT IS MEMORIZED TO THE DEVICE INNER MEMORY WHEN CLR IS CONTINUSOULY PRESSED IT IS GONE OVER THE NEXT FUNCTION NUMBER AFTER THE INPU...

Page 8: ...E WEIGHT IS CONTINUOUSLY SHANGED EVEN IN THE SMALL VALUE FOR ANY RERASONS THERE MOTION BAND F04 1 0 3 LARGE AND IT IS SMALL IN CASE THE SETTING UP VALUE IS LARGE THIS COMPENSATES THE MOMENTARY MOTION AND THUS IF THE SETTING UP VALUE IS SMALL THE MOTION IS AUTO ZERO F05 00 00 99 IN CASE THERE IS THE REST AND F 05 VALUE TO BE 10 THE ZERO POINT AFTER WEIGHING THE WEIGHTIS AUTOMATICALLY WORKED UNTIL 1...

Page 9: ... POINT KEY OPERATION MODE F08 0 STABLE VESSEL KEY 1 NON STABLE VESSEL KEY MAKING THE ZERO KEY TO PERFORM THE STABLE NON STABLE OPERATION HOLD MODE F16 0 PEAK HOLD SAMPLE MODE 1 SAMPLE MODE HOLD INPUT MODE F17 0 ON OFF PULSE TYPE 1 LEVEL TYPE SAMPLE HOLD MODE EXCEPT FOR PEAK HOLD F18 0 MANUAL STABLE NON STABLE OPERATION 1 AUTOMATIC STABLE OPERATION ...

Page 10: ... NOT REGARD THE DECIMAL POINT AND THEN PRESS SET CAL KEYAFTER THE INPUT NOTE 1 IF THE CONSTANT IS AMENDED AT RANDOM THE WEIGHT BECOMES FALSE AND THUS IT IS NOT AMENDED AT RANDOM AND WHEN THE CONSTANT IS REMOVED IT IS INPUT 2 PERFORM THE MEMO OF CONSTANT VALUE INDICATED IN THE ADJUSTMENT OF WEIGH ZERO WEIGHT ADJUSTING DIP BASIC ZERO 30000 4000 ZERO SPAN ADJUSTING DIP S W 1 2 3 4 5 6 7 8 ZERO SPAN W...

Page 11: ...ALE IS 5000 4 STEP ENTER THE INPUT OF WEIGHT OF SCALE FOR SPAN ADJUSTING SPAN HOWEVER BE SURE TO SET UP THE VALUE WHICH IS SIMILAR TO TOTAL WEIGHT OR IS MORE THAN 80 OF TOTAL WEIGHT AT LEAST 5 STEP SHOWING UP IN THE INDICATOR UP PUT THE SPAN ADJUSTING SCALE ON LOAD CELL 6 STEP SPAN ADJUSTING CONSTANT CALCULATION IT CALCULATION THE CONSTANT AND SHOWS THE CONSTANT VALUE ON THE INDICATOR IF THIS PRIC...

Page 12: ...PAN IS LARGE TURN ON SPAN ADJUSTING DIP S W 7 OR 8 ERR 06 THE SPAN IS SMALL TURN ON SPAN ADJUSTING DIP S W 7 OR 8 ERR 07 IT IS OUT OF THE SCOPE OF DATA IN SETUP IF THE ERRO OCCURS IT INDICATES THE OCCURRENCE NUMBER OF EACH ERRORFOR 3 SECONDS AND THEN IS REVERTED TO THE STATE BEFORE THE OCCURRENCE OF ERROR AND THEREFORE IS INPUT AGAIN NOTE IN CASE THE CALIBRATION IS STOPPED BY THE POWER FAILURE OR ...

Reviews: