background image

Operation:   

NOTE:  Follow the initial setup procedure on page 4 before initial operation.

1.   Plug the centrifuge into an approved 115 Volt A.C., 60 Hz. outlet.

2.   Turn the latch counter-clockwise and open the lid.

3.   Insert cushions (if needed) into the tube holders for the tube size you are  

 

    using. Refer to ‘Tube Holder Configurations’ (page 6) for assistance.
4.   Place the test tube samples into the tube holders. Be sure to follow the   

 

      rules for balanced loads.  

5.   Close the lid and turn the lid knob clockwise to its complete stop position.    
6.   Turn on the machine by turning the timer to the desired run time.
7.   The centrifuge should begin to spin.
8.  Once the timer reaches zero (0), power will be cut to the motor and the   

 

 

rotor will coast to a stop. Do not open the lid until the rotor has come to a  

 

 

complete stop.

9.   Turn the lid knob counter–clockwise and open the lid.
10.  Remove the samples. 
11.  The centrifuge is immediately ready for operation.

Page 5

Rotor Removal and Installation:

To remove the rotor:

CAUTION:  

Unplug the centrifuge from the electrical outlet at this time to 

eliminate the possibility of electrical shock or other injury.

1.   Open the lid.

2.   Remove the test tube holders.
3.   Remove the knob or nut in the center of the rotor by turning it

    counter-clockwise. A nut driver may be required.

4.   The rotor is sitting on a cone-shaped adapter. Pull the rotor up and off 
        of this adapter.

To install the rotor:

1.   Place the rotor back onto the cone-shaped adapter. You may need to 
        turn the rotor slightly to line it up properly.
2.   The rotor should slide onto the rotor cone freely.
3.   Once a proper fit has been achieved, replace the rotor knob or nut and 

        turn it until it is hand-tight.

4.   Replace the tube holders and verify that they are seated properly.
5.  It is recommended that the initial setup procedures be performed to  

 

 

        ensure that the rotor has been installed correctly and that no damage 
        has been done to the centrifuge during either the rotor installation or 

        possible rotor chamber cleaning. See page 4 for this procedure.

Safety:

Lid Safety Switch:  

The lid is secured to the top of the cabinet by a latching knob 

and pawl system. When the knob is rotated clockwise, the pawl grips the underside 
of the cabinet opening and prevents the lid from opening. A mechanical stop 
positions the pawl and prevents it from rotating completely. When rotated to the 
stop position, the pawl makes contact with a micro–switch mounted underneath the 
cabinet top. The lid safety switch prevents the centrifuge from operating while the 
lid is open. 

 

Circuit Breaker:  

The Model 614B is protected with a 4 Amp circuit breaker located 

at the rear of the machine mounted to the base. Any electrical short circuit will 
cause the breaker to cut power to the machine.

Calibration and Earth Ground Testing:

It is recommended that the top speed, ground continuity and line leakage be tested 
every two years for continued safe operation. Contact The Drucker Company for 

further information or testing availability.

The Model 614B complies with all requirements of UL standard 

61010A-1, 61010A-2-20; Can/CSA C22.2 No’s 1010.1; 1010.2.20.

Page 8

1. Problem: The rotor does not spin freely.

Solutions: – Make sure nothing has fallen into the rotor chamber.

– If there is nothing obstructing the rotor, contact The Drucker  

    

   Company for further assistance.

2. Problem: Excessive noise when the machine is running.

Solutions: – Check to see that the load is balanced.

– Make sure that nothing has fallen into the rotor chamber.

– Make sure that the nut in the center of the rotor is tight.

– Have a technician test the motor and replace it if necessary.

3. Problem: The centrifuge does not run.

Solutions: – Check the electrical outlet.

– Make sure the lid latch is turned completely clockwise to its stop

   position.

– Check the circuit breaker switch at the bottom left of the machine.  

   If the switch is white, the breaker has tripped. Contact The

   Drucker Company for further assistance.

Troubleshooting:

NOTE:

The latch must be turned completely 

clockwise to its stop position in order 

for the centrifuge to operate.

6.  

Tube Holder Replacement:

  It is recommended that the tube holders be  

 

    replaced after 24 months of use.

7.    

Remove Accessories Before Moving:

  All tube holders, samples, and caps         

       must be removed from the rotor chamber before transporting or storing 

       the centrifuge to prevent damage and injury.

(Continued)

Summary of Contents for 614B

Page 1: ...Operator s Manual Model 614B Laboratory Centrifuge Rev C P N 7711001 Reliability in Laboratory Centrifuges Since 1932 ...

Page 2: ...f both the operator and service personnel care should be taken when using this centrifuge if handling substances that are known to be toxic radioactive or contaminated with pathogenic microorganisms When Risk Group II materials are used as identified in the World Health Organization Laboratory Bio Safety Manual a Bio Seal should be employed The Bio Seal accessory for the model 614 tube holders is ...

Page 3: ...Maximum capacity 90 mL 6 x 15 mL Overall Dimensions H x W x D 8 5 in x 11 in x 12 5 in Centrifuge Motor 1 30 HP AC Nominal Acceleration Time 45 seconds Protection Breaker 4 Amp re settable Timer mechanical 1 to 30 minutes accuracy 10 Current Requirement 1 0 Amps Voltage Requirement 115 10 Volts Frequency 60 Hz Weight 18 5 lbs Before using any cleaning or decontamination methods except those recomm...

Page 4: ...s be properly grounded 2 Turn the latch counter clockwise and open the lid 3 Spin the rotor by hand check for free and level rotation 4 Close the lid Rotate the lid knob clockwise to its complete stop position 5 Turn the centrifuge on by turning the timer to 10 minutes 6 Listen to the centrifuge A smooth whirring sound should be heard After the centrifuge has passed this procedure it is ready for ...

Page 5: ... the top of the cabinet by a latching knob and pawl system When the knob is rotated clockwise the pawl grips the underside of the cabinet opening and prevents the lid from opening A mechanical stop positions the pawl and prevents it from rotating completely When rotated to the stop position the pawl makes contact with a micro switch mounted underneath the cabinet top The lid safety switch prevents...

Page 6: ... Model 614B uses a brushless A C motor It should not need servicig for the life of the centrifuge The electrical components are selected for high reliability and should not need service 5 Keep the Rotor Chamber Clean Every six months or whenever there is a tube breakage refer to the note in 3 it may be necessary to remove the rotor and clean the rotor chamber Follow the instructions on page 5 to r...

Reviews: