Drucker Diagnostics Owl Manor HORIZON 24 Flex Operator'S Manual Download Page 3

3

For the safety of both the operator and service personnel, care should be taken when 

using this centrifuge if handling substances that are known to be toxic, radioactive or 

contaminated with pathogenic microorganisms. When Risk Group II materials are used (as 

identified in the World Health Organization “Laboratory Bio-Safety Manual”), a Bio-Seal 

should be employed. In the event that materials of a higher risk group are being used, 

more than one level of protection must be provided. The use of flammable or explosive 

materials as well as those materials which have a vigorous chemical reaction is prohibited.

Supplied Accessories (Standard)

See next page for optional rotor/accessories

Model Description

HORIZON is a versatile line of centrifuges designed with 3 

settings to process Chemistry, Coag or Platelet Poor Plasma, 

and Urine specimens in the same unit. The maximum g-force 

of 2,000 xg makes HORIZON compatible with most brands 

of tubes. Cycle settings can be changed to accommodate 

custom setting.

WARRANTY:

Drucker Diagnostics warranties that this centrifuge is free from

defects in workmanship and parts for 2 years.

See the next page for a description of optional accessories.

Also included (not shown):

•  One (1) 6´ Line Cord 
•  One (1) Operator’s Manual

Intended Use

General purpose laboratory centrifuge for sample separation.

WARNING

One (1) 4-Bucket

Horizontal Rotor*

p/n 7786012

Four (4) 1.5" Green Buckets**

p/n 7713107

**for use with Owl Manor 60 mL Products

HORIZON 24 Flex, Owl Manor (00-384-092-000)

Two (2) 1.5” Green Buckets**

p/n 7713107

**for use with Owl Manor 60 mL Products

Two (2) 1.5" Purple Buckets***

p/n 7713037

***for use with Owl Manor 30 mL Products, 

HORIZON 24 Flex, Owl Manor SA (00-384-092-001)

*The horizontal rotors and rotor 

accessories are rated for a rotational 

speed of 3500 RPM (a force of 3200 xg).

200 Shady Lane, Suite 170 – Philipsburg, PA 16866, USA

+1-877-231-3115 (U.S. only) - +1-814-692-7661

[email protected] | druckerdiagnostics.com

THIS IS A UNIVERSAL CENTRIFUGE, COMPATIBLE WITH VOLTAGE 
REQUIREMENTS IN ALL COUNTRIES. THE LINE CORD WILL VARY BY COUNTRY.

Summary of Contents for Owl Manor HORIZON 24 Flex

Page 1: ...Laboratory Centrifuge Model HORIZON 24 Flex Operator s Manual P N 03 0 0002 0143 Rev C MODEL HORIZON 24 FLEX...

Page 2: ...7 Operation 7 Guide to Balanced Centrifuge Loads 8 Control Panel Functions Time Adjustment and Timer Operation 9 Speed Adjustment 9 Starting and Stopping a Run 9 Unlocking the Lid 9 Imbalance Detectio...

Page 3: ...f tubes Cycle settings can be changed to accommodate custom setting WARRANTY Drucker Diagnostics warranties that this centrifuge is free from defects in workmanship and parts for 2 years See the next...

Page 4: ...minutes 1 Weight 39 lbs Ambient Temperature During Operation 16 o C 35 o C Maximum Relative Air Humidity 90 Power Watts 280 Voltage Volts 95 to 235 Frequency Hz 50 60 Specifications General Specifica...

Page 5: ...ol Panel see below Cabinet Up Down RPM Change Buttons Running Indicator Light Latched Indicator Light Unlocked Indicator Light RPM Display Open Stop Button Start Button Up Down Time Change Buttons Tim...

Page 6: ...atched properly The centrifuge will not run unless the lid is latched properly and the LATCHED indicator light is illuminated 8 Use the up arrow button next to the RPM display to set the speed to full...

Page 7: ...nd the rotor will decelerate to a complete stop 8 The UNLOCKED indicator light will illuminate the lid light will flash and the locking mechanism will disengage allowing entry into the rotor chamber I...

Page 8: ...r a device and a counterbalance Please refer to device I F U for proper use of the counterbalance 1 device Load the opposing bucket with a counter balance 2 devices Place the two devices in opposite b...

Page 9: ...lid switch closed press the START button to begin a run Press the OPEN STOP button at any time during a run to terminate it Unlocking the Lid The lid is unlocked whenever the red UNLOCKED indicator li...

Page 10: ...ly remove the bucket Properly dispose of the sample and tube fragments and thoroughly clean both the inside and outside of the bucket Insert a new tube cushion if necessary and replace the bucket in t...

Page 11: ...ERROR SPEED The centrifuge cannot reach full speed due to a problem with the rotor an inadequate power supply or other electrical problems Press the OPEN STOP button to cancel the error and then check...

Page 12: ...l remain locked If the knob is accidentally moved and this situation should occur rotate the knob fully clockwise to its stop position and the centrifuge will resume operation Emergency Rotor Chamber...

Page 13: ...nd recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment For more...