Page 3: ...issued to protect the user and the equipment from injury or damage Use qualified personnel and good engineering practice for all procedures in this publication Maintenance The sensor must be maintained using the manufacturer s procedures and these should be carried out by authorized service agents or the manufacturer s service departments Druck com For technical advice contact the manufacturer WAR...

Page 4: ...ired the extraction and use of natural resources for its production It may contain hazardous substances that could impact health and the environment In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources we encourage you to use the appropriate take back systems Those systems will reuse or recycle most of the materials of you...

Page 5: ...on 4 3 1 General Requirements 4 3 2 Safety Measures 4 3 3 Connecting to a Pressure Source 5 3 3 1 Media Compatibility 6 3 3 2 Pressure Containment 7 3 4 Power Requirements 8 3 5 Maintenance 9 3 5 1 Visual Inspection 9 3 5 2 Cleaning 9 3 6 Returned Goods Procedure 9 3 6 1 Safety Precautions 9 3 6 2 Important Notice 9 3 7 Electromagnetic Compatibility 9 3 7 1 Power Supply and Metering 9 3 7 2 Cable ...

Page 6: ...Copyright 2022 Baker Hughes Company iv UNIK 5 00 Series English ...

Page 7: ...tong International Industrial Park No 8 Xihu Road Wujin High Tech Industrial Zone Changzhou Jiangsu 213164 China Each sensor is marked to indicate the country of manufacture 2 Description 2 1 Purpose UNIK pressure sensors hereinafter referred to as sensors are designed for continuous measurement and conversion of gauge atmospheric absolute and differential pressure into an analogue output signal o...

Page 8: ...ss to metal seal for electrical connections and a fluid filled cavity containing a silicon diaphragm with diffusion implanted resistors arranged in the form of a Wheatstone bridge The operating principle of the pressure sensor is based on the piezo resistive effect a change in resistance when applying pressure When pressure is applied the silicon diaphragm flexes changing the resistance of the imp...

Page 9: ...ccuracy description Industrial Improved Premium 5 Model number 6 Serial number 7 Pressure range limits and unit of measurement 8 Input supply voltage range and current limits Note The symbol indicates the use of direct current 9 Output voltage or current range 10 Ambient temperature range 11 Reserved for applicable certification markings Position on sensor may vary 12 Manufacturer s name and addre...

Page 10: ...ssure of the medium to be measured must be done after the shutoff valve is closed from the process and the pressure in the working chamber is made equal to atmospheric The connecting pipes must have a one way slope not less than 1 10 from the pressure collection point up to the sensor if the medium to be measured is gas and down to the sensor if the medium is liquid If this is not possible when me...

Page 11: ...impact shock mechanical and thermal stresses Do not install the equipment where it can be damaged by a material that causes corrosion Provide additional protection for the equipment if it may be damaged in service When installing power supply and signal wiring the possibility of condensate entering the sensor cable entry should be avoided 3 3 Connecting to a Pressure Source When mounting the senso...

Page 12: ...ical connector option 3 immersion fluid must be compatible with polyurethane For electrical connector option 4 immersion fluid must be compatible with Hytrel For electrical connector option N immersion fluid must be compatible with Kynar 5600 0 200 bar 0 2900 psi Fluids compatible with stainless steel 316L and Hastelloy C276 for differential versions only 201 500 bar 2915 7250 psi Liquids and grou...

Page 13: ...rsion fluid must be compatible with nitrile butadiene rubber 5800 5900 0 200 bar 0 2900 psi Fluids compatible with stainless steel 316L and Hastelloy C276 201 500 bar 2915 7250 psi Liquids and group II gases compatible with stainless steel 316L 501 700 bar 7265 10150 psi Group II liquids and group II gases compatible with stainless steel 316L Table 2 Pressure Containment Product Pressure Range Pre...

Page 14: ...DCR Passive 2 5 to 12 Vdc PDCR Linearized 7 to 12 Vdc 5600 5700 PTX 7 to 32 Vdc 5800 5900 PMP Basic 7 to 32 Vdc 12 to 32 Vdc for 0 to 10 V Output PMP Basic Configurable Maximum Output 1 V to 32 Vdc PTX 7 to 32 Vdc PDCR Passive 2 5 to 12 Vdc PDCR Linearized 7 to 12 Vdc Table 4 Supply Consumption Product Output Type Current Consumption 5000 PMP Basic 3 mA PMP Ratiometric 3 mA PMP Basic Configurable ...

Page 15: ...n contact with hazardous or toxic materials Please supply the applicable Control of Substances Hazardous to Health COSHH or Material Safety Data Sheet MSDS references and precautions 3 6 2 Important Notice Service or calibration by unauthorized sources will affect the warranty and may not guarantee further performance If the equipment has hazardous area approval the approval will also be invalid 3...

Page 16: ...t through which it is going to run Screened cable should always be used where electrical noise is present Good cabling practice will be reflected in signal quality 3 7 3 Earthing Grounding For the screening of the cable to be effective it is essential that the screen or drain conductor is permanently bonded to earth ground This should take place at the monitoring end of the cable as close to the p...