background image

K0455 Revision A

Preparation

To use this charger, make sure you have 

these items:
• IO620-PSU: 5 V DC power supply. A 

standard item with the DPI

 

620G 

calibrator.
Alternative: A USB mini Type B cable 

(IO620-USB-PC); not a standard item.

• IO620-BATTERY: Li-ion battery.

Assembly and Operation

In a dry location, put the charger on a flat 

surface near an applicable power socket: AC 

wall socket or USB computer socket. Then:
1. Connect the power supply to the charger: 

(a) 5 V DC power supply OR 

 

(b) USB mini Type B

2. Put the battery in the charger until the 

guides (c), (d) and (e) hold it in position.

Specification

Charge Temperature: 

0 to 40°C (32 to 

104°F). When the charger senses the 

temperature is outside this range, it stops 

charging.

Storage Temperature: 

-20 to 70°C (-4 to 

158°F).

Humidity: 

0 to 90% relative humidity (RH) 

non-condensing.

EMC: 

BS

 

EN

 

61326-1:2013

Electrical Safety: 

BS

 

EN

 

61010:2010

Size (L:W:H); weight: 

 97 x 83 x 22 mm 

(3.8

 

x 3.3 x 0.7 in); 

 50 g (1.8 oz)

Customer Service

Visit our web site: Druck.com

Copyright 2008 Baker Hughes Company. This material contains one or more 

registered trademarks of Baker Hughes Company and its subsidiaries in one or 

more countries. All third-party product and company names are trademarks of their 

respective holders.

Charge Method

Charge Time (to full 

capacity)

5 V DC power supply

 6.5 hours

USB mini

 

Type B 

cable

 13 hours (with a 

500

 

mA supply)

LED Indications

Condition

Flashing red and green 

LEDs

No battery

Continuous red LED

Charging

Continuous green LED

Charging 

complete

B

1

2

OR

a

b

e

c

c

Red LED

Green 

LED

d

4

5

6

7

Reviews: