Copyright 2014 Baker Hughes Company.
10 | ADTS Safety and Installation Guide–English
1. Clean the external surfaces of the unit using lint-free cloth and a soft brush. Remove
stubborn dirt using a damp cloth and mild detergent and allow to dry.
2. Clean the electrical connector using a soft brush. Remove stubborn dirt using a damp cloth
and mild detergent and allow to dry.
4. Specification
4.1 Operating Environment
WARNING
This equipment is not rated for use in potentially explosive
atmospheres.
Storage temperatures
ADTS542F: -20°C to 70°C (-4° to 158°F)
ADTS552F )
ADTS553F ) -20°C to 70°C (-4° to 158°F)
ADTS554F )
Operating
temperatures/environment
ADTS542F: -5°C to 50°C (32° to 122°F)
ADTS552F )
ADTS553F ) -5°C to 50°C (32° to 122°F)
ADTS554F )
Operating humidity: 5% to 95% RH (condensing)
Indoor & Outdoor (non-continuously) use
ADTS Touch universal power
supply
Operating temperature: 10°C to 50°C (50° to 122°F)
Operating humidity: 5% to 95% RH (non-condensing)
Power supply
Universal input Mains Power Supply.
Class 1 equipment requiring a protective earth connection.
TXAH 250V HRC Time-delay (T) 250Va.c. rated fuses (X = amp
rating of fuse).
Overvoltage category II, Pollution degree 2
The supplied mains/PSU cable protective earth/ground conductor
must be connected to the power supply protective earth/ground
system.
Electrical safety
EN
61010-1, UL
61010-1, CSA 22.2, No. 61010-1 and IEC
61010-1
Operating altitude
Maximum 7500ft (2300 metres)
Ingress protection
ADTS542/552/553/554 IP33, ADTS Touch IP65 (EN60529)