Example:
The below screen shows the scale value changed from the default value of
0.0100 to 3.0000 for mbar.
•
To set a negative value, press the
key until the UP / DOWN arrows flash, then
press the
key to toggle the negative value symbol.
•
To move the decimal point, press the
key until the decimal point flashes, then
press the
key to move the decimal point to the right one digit at a time. If it is taken
too far press the
button and repeat the process.
•
The scale factor that is set for a unit is automatically applied for all other units. To verify
this, press the
key and navigate to other units. The scale factor values updates.
For example, after setting a scale factor of 3.000 for mbar, a scale factor of 0.0030 for
bar is shown and so on.
7. Press the
key to return to the sensor reading screen.
8. To apply the custom unit to the pressure reading:
•
Navigate to the sensor reading screen as shown in step 1 above.
•
Press the
key until a down arrow is visible at the bottom as shown below. The
custom unit is now applied.
Note:
The down arrow means the custom unit is selected.
•
The reading which is visible will be scaled to the scale factor that is configured.
3.2.9 Software Version
This setting shows the application software version on the instrument.
To see the application software version:
1. Ensure the display is on the sensor reading screen.