160 | DPI610E Instruction Manual–English
Chapter 9. Data Log
9.3 Periodic Trigger Options
The options
TIME INTERVAL
,
LOG DURATION
and
DATA POINTS
are only available for use
when
PERIODIC
is the trigger mode for recording data results.
9.3.1 TIME INTERVAL
When a Periodic trigger is selected, the option to set the Periodic interval is available from the
Data Log Setup
screen. The interval determines how long each data point is logged and it is
expressed as HH:MM:SS.S. The supported time interval range is between 00:00:00.2 to
23:59:59.9. The minimum interval supported is dependent on the functions configured in the
Calibrator channels. See table below.
9.3.1.1 How to set the TIME INTERVAL
Function
Minimum Interval (HH:MM:SS.S)
Internal Barometer
00:00:05.0
Internal Pressure
00:00:00.2
External Sensor (Pressure and RTD)
00:00:00.2
Current / Voltage / Millivolt Measure
00:00:00.5
Current Source
00:00:01.0
HART
00:00:00.5
Step 1
Select
TIME INTERVAL
from the
Data Log
Setup
screen.