Copyright 1996 Baker Hughes Company.
English–ADTS 403/405 Calibration Manual | 27
Function and Comments
[EXTERNAL]
[Ps & Pt (abs)]
data entry
[AIM]
[RATE]
Go to new pressure set-point.
Set Rate set-point.
Turn controllers OFF.
data entry
[OFF]
[ON]
[END Ps]
Turn controllers ON.
Exit Ps check, continue with Pt check.
[Ps]
data entry
[AIM]
[RATE]
data entry
[ON]
[OFF]
Go to new pressure set-point.
Set Rate set-point.
Turn controllers ON.
Turn controllers OFF.
[Pt (abs)]
data entry
[AIM]
[RATE]
data entry
[ON]
[OFF]
Go to new pressure set-point.
Set Rate set-point.
Turn controllers ON.
Turn controllers OFF.
[Ps & Pt (abs)]
data entry
[AIM]
[RATE CALIBRATION]
[RATE]
data entry
data entry
[ON]
[OFF]
Go to new pressure set-point.
Set Rate set-point.
Turn controllers OFF.
Turn controllers ON.
[000]
Set Service (Calibration) PIN.
Key/selection
Function and Comments
[END Ps]
Exit Ps check, continue with Pt check.