7) Maintenance Schedule & Daily Checklist
All Drive DeVilbiss products are designed for minimum maintenance, however some safety
checks and procedures are required. A schedule of DAILY tasks are detailed below. Daily
checks and a yearly service, inspection and test will ensure a lift is kept in optimum safe
working condition. A list of spare parts are available upon request.
DAILY CHECK LIST: Drive DeVilbiss Healthcare recommends regular inspection and
maintenance. Please refer to the chart below.
Check item
Detail
Initial
Before
operation
Annual
service
Sling
1. Examine wear or fraying, if any, do not use
v
v
v
Spread bar
1. Check the spreader bar for rotating and swing
freely. Check for wear on the cradle pivot,
make sure all attachment is firmly fixed on the
boom.
2. Make sure clip of hook is working properly.
v
v
v
Mast
1. Check the mast is engaged into base socket and
fixed properly.
2. Check actuator is well installed on mast with bolt
and nut fastened.
v
v
v
Boom
1. Check bolt of joint with mast is properly fixed.
2. Make sure there is only slightly side movement of
the boom.
3. Make sure boom is able to rotate smoothly and
freely.
4. Check actuator is well installed on boom with bolt
and nut fastened.
v
v
v
Actuator
1. Actuator is well fixed.
2. No abnormal noise is heard
3. No descending when with loading except
handset down button is touched.
4. Check manual lowering function is well working.
v
v
v
Control box
1. Check emergency stop button function.
2. Check handset connecting is properly.
3. Check actuator connecting is properly.
4. Check up-down function is properly.
v
v
v
Summary of Contents for DeVilbiss Healthcare Gravis FLP500
Page 17: ...17 ...
Page 18: ...18 ...
Page 23: ...23 8 Contact detail Communication and service detail Drive Medical ...
Page 27: ...27 12 Commented SM2 EMI is FPO only Angelo Aidan to supply Commented SM3R2 ...
Page 28: ...28 Commented SM4 EMI is FPO only Angelo Aidan to supply ...
Page 29: ...29 Commented SM5 EMI is FPO only Angelo Aidan to supply ...