background image

It is extremely important that the lift be inspected 

before each use. The following items need to be 

checked:

a

. All nuts and bolts should be tight.

b

.  The lifter should move freely with caster  

brakes off.

c

.  Check the brakes on the rear casters for proper 

operation.

d

.  Make sure base adjustment mechanism is 

operating easily.

e

.  Double check to ensure the “U” bracket at the end 

of the boom and the “S” hook on the hanging bar 

are secured tightly.

f

.   Work the Actuator to ensure that the battery still 

has enough power. When the battery is 50% below 

full capacity, it will give a “beep” warning and will 

need to be re-charged immediately.

g

.  Before initial use, place a drop of oil on the 

following areas:

1

.   Where mast and boom connect, hanging bar hook, 

pump handle hinge, and caster axles.

2

.

  

Where pump and boom attach.

3

.

  

Lubricate above mentioned every 2 to 3 months.

d

.  Hook the hanging strips with the hanging bar. If a 

chain is used, make sure the “S” hooks are away 

from the patient.

e

.  Lift the patient by pressing the UP button on the 

Control Handset.  DO NOT lock the casters of 

the Patient lift when lifting an individual.  Casters 

MUST be left unlocked to allow Patient Lift to 

stabilize during the lifting procedure.

f

.   Lift patient until his/her feet will swing easily off 

the bed keeping patient facing the attendant.

g

.  Transfer patient to the wheelchair or commode and 

position above seat surface.  Lock brakes of both 

lifter and commode or wheelchair.

h

.  Press the down button on control handset to 

gradually lower the patient.

i

.   During the descent, assist patient to attain correct 

sitting posture.

j

.   For transferring to wheelchair, when patient is  

seated, push down on boom to slacken hanging 

strips of sling. Patient can remain seated in sling.

k

.  For commode use, adjust clothing before moving 

lift to straddle commode. Keep hanging strips taut 

and make sure patient is in a comfortable position.

how to operate the 

electric patient lift: 

1

.   Spread  the  base legs to the widest position be-

fore lifting.  

2

.  Keep the patient centered between the base legs 

and have the patient face the attendant.

3

.  Be extremely cautious and use restraint straps for 

spastic or severely handicapped patients.

4

. How to use Standard or Commode Slings.

a

.  For smooth and easy lifting, have the lift, chains, 

sling, commode or wheelchair in ready position.

b

.  With the patient in the center of the bed, roll 

patient on side away from the attendant.

c

.  Roll the patient on side toward the attendant and  

center the patient on sling. With the base of lifter 

under bed, press the down button on the control 

handset to lower the boom (#11).

1

.   Patient lifter is a transfer device that allows 

patient attendant to transfer patients from bed to 

wheelchair or commode...etc. It should not be used 

to transport patients.

2

.  When lifting the patient, make sure the base legs 

are in the most widely opened position.  DO NOT 

lock the casters of the Patient lift when lifting 

an individual.  Casters MUST be left unlocked to 

allow Patient Lift to stabilize during the lifting 

procedure.

3

.  When battery power goes below 50%, the 

controller has an audible “beep” warning. When 

the low battery warning is heard, charge the 

battery immediately. Please note that a lifted 

patient cannot be lowered down when battery is 

out of power.

4

.  Push the “emergency” red button if the control 

unit system is not functioning properly.

important safety instructions: