background image

Indications For Use:
Aerosol delivery to manage the symptoms of patients 
with  Asthma,  Chronic  Lung  Disease  and  Cystic 
Fibrosis.

CAUTION:

Federal law restricts this device for sale or use on the 
order of a physician.
To reduce risk of electrocution:
- Always unplug unit immediately after use.
- Do not use while bathing.
- Do not place or store the unit where it can fall or be 
pulled into a tub or sink.
- Do not place or drop into water or other liquid.
-  Do  not  reach  for  a  product  that  has  fallen  into 
water. Unplug immediately.

WARNING:

To reduce risk of electrocution, fire or injury:
- Electrical shock hazard. Do not remove the cabinet 
or open the cover.
- Disconnect the power cord from the electrical outlet 
before cleaning or servicing.
- Do not place this equipment near hot, sparking or 
burning objects.
- Do not use oil or grease on or near this device.
- Turn the unit off when it is not in use.
- Keep the cord away from heated or hot surfaces.
- Never drop or insert any object into any opening.
- Never block the air opening of the product or place 
it on a soft surface, such as a bed or couch where the 
opening may be blocked.
- Avoid the operation in wet or damp locations.
- Unplug the unit before filling the nebulizer.

 Electrical Requirements:  

115VAC/60HZ

 Power Consumption: 

90W

 Dimensions: 

13.7” (H) x 7.5” (W)  

 

x 4.7” (D)

 Weight: 

6.4 lbs.

 Maximum Pressure: 

30 PSI

 Operating Pressure: 

12-14 PSI

 Liter Flow Range: 

9 LPM  

 Particle Size:                      Less than 5 microns

 Operating Temperature:  50°-104°F.

 

NOTE: 

Conditions  may  vary  based  on  altitude,  

barometric pressure and temperature.

1. Place compressor on a stable, sturdy and flat sur-
face such that the unit can be easily reached when 
you are seated.
2. Pull out the power cord from the base of the unit 
and make sure that the unit is in the “OFF” (O) posi-
tion by pressing on the left side of the switch.
3.  Plug  the  power  cord  into  a  115  VAC/60  Hz  wall 
outlet.
4. Connect one end of the nebulizer tubing into the 
air outlet connector.
5. Add the prescribed medication through the open-
ing  on  cap  using  an  eyedropper  or  pre-measured 
dose container.
6. Assemble mouthpiece and T-piece and insert into 
the  top  of  the  nebulizer  cap.  If  an  aerosol  mask  is 
used, connect the bottom of the mask to the top of 
the nebulizer cap.
7.  Connect  the  tubing  to  nebulizer  air  inlet  
connector.

important safeguards:

specifications:

operating instructions:

Summary of Contents for Power Neb II

Page 1: ...ses ann es Dans l ventualit d un d faut couvert par cette garan tie nous r parerons ou remplacerons le mat riel d fectueux notre discr tion Cette garantie ne s applique pas des probl mes caus s par un...

Page 2: ...bed or couch where the opening may be blocked Avoid the operation in wet or damp locations Unplug the unit before filling the nebulizer Electrical Requirements 115VAC 60HZ Power Consumption 90W Dimens...

Page 3: ...cap for one half hour 5 Remove the components from the solution an let dry 1 The filter of air compressor should be replaced when it turns grey Additional filters can be purchased from the dealer or p...

Page 4: ...dos Desenchufe la unidad antes de llenar el nebuliza dor Requerimientos El ctricos 115VAC 60HZ Consumo de Energ a 90 W Dimensiones 34 8 cm Alto x19 05 cm Ancho x 11 9 cm Largo Peso 2 9 Kg Presi n M xi...

Page 5: ...en agua caliente con jab n para platos Enjuage completamente 4 Mezcle una parte de vinagre blanco con tres partes de agua destilada en un envase limpio y sela como soluci n desinfectante Sumerja la p...

Page 6: ...areil D connecter l appareil quand il n est pas en fonc tionnement indications d emploi mise en garde Ne jamais laisser le cordon d alimentation toucher une surface chaude ou chauffante Ne jamais lais...

Page 7: ...l embout buccal 10 Si un masque a rosol est utilis couvrir la bouche et le nez 1 Le filtre du compresseur d air doit tre remplac quand il devient gris Les filtres de rechange doivent tre achet s chez...

Page 8: ...esgaste normal y necesitan reemplazo peri dico Si usted tiene alguna pregunta acerca de su aparato Drive o esta garant a por favor contacte a un repre sentante autorizado de Drive Your Drive brand pro...

Reviews: