1- Clean the outer surfaces with a wet cloth and dry immediately. The jet
nebulizer should be replaced if obstructed by any medicine residue or
other substances.
WarnInG:
- Jet nebulizer and accessories should not be boiled.
- Do not let the unit get wet nor submerge it into any liquid.
- Do not use any abrasive substances to clean the unit.
2- The air filter is to be replaced once a month or earlier, if it became gray.
aTTenTIOn:
- use only the manufacturer’s original filters.
- Do not operate the unit without an air filter.
ImPOrTanT:
- If medicine enters the compressor area, clean up the device with a dry
cloth and stand it idle for 5 minutes by placing it upside down. after 5
minutes, operate the device for 2 minutes without the nebulizer.
- The unit should not be opened for any reason. The parts contained cannot
be repaired by any unauthorized person; the compressor requires neither
lubrication nor maintenance.
- after cleaning operations, cover the nebulizer with the unit lid.
WarnInG: in order to avoid any infection risk due to contaminated
medicine, it is recommended to clean up parts after each use.
nebulize only with products prescribed by a physician and follow the
instructions. Do not use any balsamic or aromatic substances. Do not
use any volatile fluids, like benzene or alcohol, etc. The life time of the
accessories, using them according to the instructions of this manual is
of 5 years.
cleaning and maintenance