USER AND MAINTENANCE MANUAL FOR

 

Integra et Selectis 

SHELL SEATS

 

4

III - SAFETY INSTRUCTIONS

The footrest (option) should always touch the floor when  
the chair is in trendelenburg (tilted forwards).

3.1) Carrying and lifting

It is totally forbidden to lift the chair when the user is seating 
on. To lift the chair alone, exclusively use the fixed frame parts. 
Never use the legrest.

3.2) Transport in a vehicle

Integra 

and

 Selectis

 chairs should never be used as a 

passenger seat in a vehicle, even if it is equipped and strapped. 
If the chair is to be transported in a vehicle, it is preferable to
anchor it with straps in order to secure it and remove the posible
accessories (shelf, etc.).

CAUTION

: the weight of the chair or of some parts may be 

so heavy, that it may be necessary to be helped for lifting.

3.3) Pinching risks

There may be some pinching risks when handling the chair or 
the accessories. The healthcare professional should warn the 
user against those risks.

3.4) Accessories and options

 

Thanks to the different options, such as the ergonomic headrest, 
and some other devices, listed in this manual, the healthcare 
professional can adapt the chair to the user’s needs. In no case 
should some other devices or options be associated to the chair 
before being sure they are compatible. Only the accessories pro-
vided by Drive DeVilviss Healthcare france should be used.

3.5) Positioning

 

(pressure of the human body onto the soft tissues)

Some users suffer from pathologies favouring bedsores when 
they are seating for long whiles. It is the healthcare professional’s 
responsibility to prescribe a chair equipped with equipment 
necessary to relieve the support points, such as a pressure-relief 
cushion.

CAUTION

: After use, always re-tension the seating pad 

upholstery.

3.6) General safety rules

Always lock the brakes when the user goes into or out from the chair.

Never go onto an escalator.
Do not smoke when sitting in the chair.

3.7) Precautions relative to electromagnetic 
compatibility

The seat was designed to meet the international electromagnetic 
compatibility requirements.
However, appliances producing significant electromagnetic fields 

may affect the seat behaviour. Should the case arise, please 
contact the distributor.

l

 

The presence of long distance transmitters and receivers, such 

as diffusion transmitters (TV and radio antenna towers), amateur 
radio stations and other devices may lead to malfunctioning. In 
case of doubt, do not use the electric shell seat anymore until the 
problem is solved.

l

 

Using the electric shell seat may lead to malfunctioning of 

some sensitive equipment. In case of doubt, do not use the shell 
seat anymore until the problem is solved.

IV - DELIVERY AND COMMISSIONING

The chair is delivered according to the prescription sheet. Any 
modification totally binds the healthcare professional.

All Drive DeVillbiss Healthcare France chairs are controlled in 
our factory and specially packaged for transport.
Before unpacking the chair, check the box has not been damaged, 
then make sure all the accessories and options are present, including 
the use instructions and the service warranty book. The accom-
panying documentation should imperatively be given to the user 
so as he/she understands the chair use and warranty conditions.
After commissioning the chair, the healthcare professional or an 
authorised person will perform the adjustments specific to the 
user. It is the professional’s responsibility to give the security and 
use instructions to the user, as well as the warnings described 
in this manual. The professional should also make sure the user 
understood these instructions. 
The packaging boxes and the different protective elements of 
the chair should be sent to an approved waste treatment centre.

V - USE RECOMMENDATIONS

This chapter describes some techniques to optimally use the 
chair. Moreover, these techniques may be useful to overcome 
everyday-life obstacles.
General and hygiene recommendations: The chair should be 
regularly cleaned according to the recommendations described 
in the following pages, and it should never be used if there any 
doubt about its proper performance.

5.1) Seat belt

  
When the user has difficulties to be seating or may lose his/her 
balance, the seat belt should be used. A healthcare professio-
nal should install and adjust the seat belt. It is important not to 
tighten too much the belt and to adjust it according to the cir-
cumstances. 
The belt fastens in front by fitting it tightly around the patient. 
The seat belt is adjusted by adjusting the strap length thanks to 
a loop; opening and closing are possible thanks to two buckles 
(male and female), one of which has a locking system.

Note: there is also an optional harness which harmonizes with 
the seat belt.

Summary of Contents for Integra

Page 1: ...Shell seats Integra Selectis USER GUIDE Plus d informations More informations More information Integra Selectis ...

Page 2: ...ENANCE 7 1 Disinfecting p 6 8 YEARLY MAINTENANCE REPAIRING WARRANTY 8 1 Main procedures p 6 8 2 Accessories p 6 8 3 Storage p 6 8 4 Chair service life and end of life p 6 8 5 Warranty conditions p 7 9 OPTIONS AND TECHNICAL INFORMATION 9 1 General presentation p 7 9 2 General terminology p 7 9 3 Options p 8 10 ASSEMBLY AND ADJUSTMENTS 10 1 Seat depth only on M version p 9 10 2 Seat height p 9 10 3 ...

Page 3: ...as abnormal play or if the upholstery ele ments are torn or undone 2 3 Outdoors use The chair can be used outdoors provided that the use is limited to terraces interior courtyards and private roads and the user should be always accompanied by someone who should take into account the road and weather conditions In case of important ramps slopes kerbs or superelevations the chair may tip over When m...

Page 4: ...tters and receivers such as diffusion transmitters TV and radio antenna towers amateur radio stations and other devices may lead to malfunctioning In case of doubt do not use the electric shell seat anymore until the problem is solved l Using the electric shell seat may lead to malfunctioning of some sensitive equipment In case of doubt do not use the shell seat anymore until the problem is solved...

Page 5: ...ying on the seat Do not lock the brakes as it is preferable to reverse rather than tip over CAUTION Never exceed the chair stability limit Never perform these manoeuvres near stairs or obstacles that could make the chair unstable 5 7 Transferring Transferring is only possible forwards After applying the brakes onto the wheels tilt the shell forwards Shell in maximum forward position The user may g...

Page 6: ...he fastenings and hinge pins with systematic repla cement of the loosened nylstop nuts and worn screws damaged thread reduced screw section and rust traces l Lubricate the hinge pins l Control of all welding and systematic replacement of the parts whose welding has cracks come off tear off the twisted out of shape or buckled parts should systematically be replaced If some paint comes off or flakes...

Page 7: ... four wheel telescopic leg lSelectis VHP wheelbase lM4 non inflammable class coloured PVC upholstery with a storage bag lShell angle from 3 forward to 26 backward lIntegra and Selectis are available in 4 versions M type Manual inclination of the shell aided with blocking Au tomatic inclination legrest M T type Manual inclination of the shell aided with blocking Electric transfer seat with remote c...

Page 8: ...on the patient s heightselected on the dimension sheet Type M 450 posi 1 490 posi 2 530 posi 3 Types MT et ET 480 posi 1 520 posi 2 560 posi 3 Types M et POSTURE 230 Types MT ET 200 Integra 780 Selectis 890 2 Type POSTURE 6 27 Type POSTURE 490 posi 2 530 posi 3 posi 1 incompatible Preset depending on the patient s height selected on the dimension sheet and modifiable by the distributor if necessar...

Page 9: ...ht or left Remove the shelf every time you want to move with the chair CAUTION Do not use the shelf edges to straighten the chair 10 4 Seating pad The standard seating pad is detachable and can be replaced by a pressure relief cushion prescribed by a specialist and compa tible with Integra and Selectis chair seat size and provi ded by Drive DeVilbiss Healthcare France 10 5 Shell tilting M M T and ...

Page 10: ...is possible to tilt the chair backwards and forwards but it is also possible to lift the seat back for easier transfer Standing up After checking the brakes are engaged straighten up the chair with the remote control until the seat is in horizontal position While still pressing the remote control button firmly hold the armrest front part rest your feet on the floor and while the seat is rising sli...

Page 11: ...lf Tools a 5 mm hexagonal key and a 13 mm spanner Insert the A shelf support into the sleeve located under the seat the tube C has to be in support on the foot position the mounting hole in front of the hole adapted to the chair dimensions Then tighten the screw and its nut to fix the assembly Join the B shelf arm in the A shelf support Thus the shelf can be positioned in position of use or retrac...

Page 12: ...ssible to quickly and simply come back to the power supply mode you just have to unplug the battery and to directly link the power supply to the chair power cord photo n 4 Photo N 4 10 13 b Precautions Regularly check the condition of the battery and cables In case of doubt immediately unplug the battery Very regularly check the box and cables do not have defect such as cut crack distortion due to...

Reviews: