9.2 Electrical Data
Voltage in:
230V±10%, ~50/60Hz.
Current
in:
1,25A
Mains
cable
fuse: 3-5A
*Duty
cycle
10%
2 mins
of continuous use followed by
18 mins
n
ot in use.
Electrical safety standards
BS EN 60601-1: 2006
EMC
BS
EN
60601-1-2:2007
The Treatment Chair range are compliant to the requirements of 60601-
1-2: 2007, the RF emissions are very low and are not likely to cause any
interference to nearby electronic equipment, however interference to sensitive
equipment
is
possible.
Electrical
shock
prevention:
Liquid ingress protection
IPX4
The electrical system is only suitable for use when:
Ambient temperature
+5° to +40°C
Humidity
20% - 90% at 30°C
Qualified users of the Treatment Chair can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters). If the Treatment Chair or any
alternative equipment is found to be operating abnormally mitigation measures
are to be taken, such as the separation distances being increased and/or the
products being re-orientated.
*Electrically operated chairs are intended to be operated intermittently rather than continuously. If the chair is
operated continuously for up to 2 minutes it must then be left for at least 18 minutes before re-use to allow the
electrical system to cool sufficiently.
If the chair is continuously used for an extended period of time and it
exceeds the duty cycle the control box may become temporarily disabled or irreparably damage.