13
Problem
Solution
No power to the unit
•
Ensure the pump is switched on at the side of the pump and not
on standby
•
Ensure the mains lead is connected to the mains and also the
pump
•
Ensure there are no blown fuses in the pump or mains lead.
Replace accordingly
Low pressure indicator is
lit
•
Ensure the correct comfort setting has been selected
•
Ensure the connections between the feed pipe and the pump unit
are secured
•
Ensure the CPR tag is engaged
•
Ensure all tubing connections along the cells are secure
•
Assess the cells for any punctures which maybe causing a leak
Patient is bottoming out
•
Pressure setting may be inadequate, adjust comfort range
No air is being produced
from some of the air
outlets of the tube
connectors
•
This is normal in alternating mode. Air outlets take turns to
produce air during their cycle time
•
If fault still continues, a Drive DeVilbiss Healthcare approved
engineer will be required to repair the fault
Pump is noisy
•
Tightly secure the bed hooks to the bed frame as this will reduce
vibration and limit noise
•
Ensure pipe work is not kinked
•
If fault still continues, a Drive DeVilbiss Healthcare approved
engineer will be required to repair the fault.