11.2 Fault Detection
This AutoFold Powerchair has an automatic fault detection
system that will alert you to any issues. When a fault occurs, the
leds on the battery gauge may light up with a series of audible
beeps. The issue can be identified using the table below. If the
fault does not appear on the table below, do not use the powerchair, turn off the power
and contact Drive DeVilbiss Healthcare Ltd.
Beep Code Battery Gauge
Resolution
Resolution
Beep
2 + 3
RHS motor is
disconnected
or in freewheel
mode
Switch off powerchair.
Check right motor is in
'Electric' drive mode.
Ensure motor cables are
plugged into the joystick
loom. Restart powerchair
Beep
2 + 5
LHS motor is
disconnected
or in freewheel
mode
Switch off powerchair.
Check right motor is in
'Electric' drive mode.
Ensure motor cables
are plugged in. Restart
powerchair.
Beep
2 + 7
Joystick is out of
neutral during
power up
Switch off powerchair.
Ensure joystick is released.
Restart powerchair.
No beep
All LEDs Ripple
Charger is
plugged into
control unit
Unplug charger from
the powerchair and then
restart.
LED :
= lit and
= not lit
11.3 Repair
Contact your provider to discuss the replacement of components on the powerchair.
Some spare parts may be replaced by the user under instruction from service
personnel, however other components will require installation by a service engineer.