ME1310 Antenna and
Propagation (3D)
Lab 1 - 16/18
S
21
Calibration
Figure 13
– Setup for S
21
calibration
1. Set up the connections as shown in Figure 13.
2. On the N9912A front panel, press
Mode
. Use the softkey to select
NA
, and then select
S
21
.
3. Press
Cal
(button #5) on the N9912A front panel. Use the softkey to select
Normalize
.
4. Then, select
Thru.
5. Select
Measure
after completing the setup.
6. Upon the completion of Step 5, press
Finish
to complete the Norminal calibration. The calibration
can be repeated again at any time in the event that the calibration kit was not properly connected
during the calibration process.