B/0 4 - 4
9 Low Inflation Pressure
Phenomenon:
During blood pressure measurement, the inflation pressure keeps to
be low.
Inspection method: Check if the NIBP cuff or extension tube is broken.
Solution:
Replace it with good NIBP cuff and extension tube.
10 Abnormal IBP Values
Phenomenon:
Displayed IBP values are different than expected.
Inspection method: Check tubing for bubbles.
Solution: Rezero
the
transducer.
11 Wrong displaying of IBP waveform
Phenomenon:
Displayed IBP waveform is not regular.
Inspection method: Check if the cable is outburst.
Solution:
Replace good quality cable.