DRE Waveline Plus
Service Manual
Page 1: ...DRE Waveline Plus Service Manual...
Page 2: ...cification 2 4 F Requirements 2 8 SECTION 3 Maintenance and Calibration 3 1 A Maintenance Schedule 3 1 B Manufacturer Responsibility 3 1 C Visual Inspection 3 1 D Cleaning 3 2 E Thorough Inspection an...
Page 3: ...eparations before use and operation principles Chapter 3 Maintenance Includes making maintenance plan visual inspection cleaning check and test Chapter 4 Troubleshooting Introduces the monitor structu...
Page 4: ...type applied part float insulation complies with IEC 60601 1 particular requirements with higher requirements for electric shock protection than B type applied part Defibrillator protection CF type a...
Page 5: ...portance Hazard indicates the potential cause which may result in injury to human body Warning indicates the potential hazard or unsafe action which may result in death or serious injury if not avoide...
Page 6: ...sy to carry advanced function easy to use 1 Dimension 320mm W x 250mm H x170mm D 2 Weight 7 95 Kg 3 Screen 10 4 4 Unity Network The device can be linked with central station B Front Panel The front pa...
Page 7: ...When a waveform is being frozen press this key to unfreeze the waveform and restore waveform sweep Press this key to enable or disable Alarm sound Press this key to activate the recorder to print Pres...
Page 8: ...950 for data processing equipment and IEC 601 1 for medical equipment Furthermore all configurations shall comply with the valid version of the system standard IEC 601 1 1 Everybody who connects addit...
Page 9: ...monitoring waveform and data Loading thermal paper Open the door of the recorder by pressing in the direction as indicated by the arrow place print paper inside according to the diagram on the recorde...
Page 10: ...Visual Input 5 lead ECG patient cable QRS Indicator Audible and Visual Alert Waveform Storage 6 minutes Sweep Speed 12 5 25 50 mm sec Gain Selection 4mV 2mV 1mV 0 5mV 0 25mV Auto Trends 2 hours 4 hour...
Page 11: ...ge Systolic Adult pediatric 30 254mmHg 4 0 33 9kPa Neonate 30 135mmHg 4 0 18 0kPa Diastolic Adult pediatric 10 220mmHg 1 3 29 3kPa Neonate 10 110mmHg 1 3 14 7kPa Cuff Pressure Range Adult pediatric 0...
Page 12: ...0 V pixels 5 Recorder Optional Type Built in 2 channel thermal array recorder Print mode Text or waveform Waveforms Real time or alarm triggered Resolution 400dpi vertical 800dpi horizontal Annotation...
Page 13: ...produce heat while working So the monitor should be located in a ventilated place Caution Do not locate the monitor in an enclosed area that may restrict the heat dissipated by it Caution The monitor...
Page 14: ...lified personnel shall perform NIBP calibration for the monitor B Manufacturer s Responsibilities If customer has not implemented the maintenance plan which results in inaccurate measurement of the mo...
Page 15: ...r 4 Do not use the following detergent abrasive detergent alcohol detergent acetone detergent E Thorough Inspection and NIBP Calibration 1 Test Equipment Multi functional simulator ECG cable 5 lead AH...
Page 16: ...quare wave is overlapped over the ECG waveform 10 ECG test is complete 4 Respiration test 1 Connect ECG cable to the ECG socket properly and set the simulator as follows respiration baseline resistanc...
Page 17: ...ug the oximetry cable Pulse oximetry test is complete 8 Non invasive Blood Pressure Test and Calibration The accuracy of the NIBP reading of the monitor is decided by NIBP calibration and the measurem...
Page 18: ...ayed on the central station monitoring system When the above tests are complete turn off all the test devices and monitor and remove the test devices away from the monitor F Electric Safety Test Elect...
Page 19: ...e No display after power on Phenomenon When the monitor is powered up there is no display on the screen and LED does not light up Inspection method Without connecting the monitor to AC power supply ch...
Page 20: ...ery humid and if the monitor has moisture inside 2 Check if the electrode s quality is good and the skin surface where the electrode is applied is clean Solution 1 Turn on the monitor and let it run f...
Page 21: ...de which will affect the measurement results Do not measure blood pressure and SpO2 on the same arm Pressing the arm will affect SpO2 measurement If SpO2 waveform channel shows No signal received it i...
Page 22: ...Solution Replace it with good NIBP cuff and extension tube 10 Abnormal IBP Values Phenomenon Displayed IBP values are different than expected Inspection method Check tubing for bubbles Solution Rezero...
Page 23: ...o the update speed of the displayed waveform on the screen There are three waveform speeds for selection 12 5mm s 25mm s 50mm s Default is 25mm s Line Mode 1 2 to select waveform described with single...
Page 24: ...display mode on the left side e g SpO2 100 90 in the above figure it is the parameter displayed in the trend graph and the alarm upper and lower limits On the right side e g 2 hours in the above figur...
Page 25: ...graph 2 hours 4 hours 8 hours 24 hours 48 hours NOTE x When press the knob to confirm the Parameter selection the cursor enters the parameter box on the right Rotate the knob to select the parameters...
Page 26: ...will stop Press the knob again the recall will continue Rotate the knob left or right you can scroll the stored ECG waveform backward and forward In the end press the MAIN MENU key on the front panel...
Page 27: ...te Knob Press Knob Triggered printing time can be set to print the waveforms starting from 5 seconds before alarm to 10s 15s 20s or 30s after alarm To set triggered printing time MAIN MENU Rotate Knob...
Page 28: ...turned on the system will display the time and date in the time status box To set up the time MAIN MENU Rotate Knob Setup Press Knob Time Rotate Knob Press Knob Rotate Knob Press Knob Rotate the knob...
Page 29: ...reen Repeat the above procedure to return to normal monitoring mode 5 6 5 Clearing Trend When monitoring a new patient first press Clear Trend to clear the trend graph of the previous patient to avoid...