Application Range

The DW-K2800 Nucleic Acid Analyzer can be used to measure:

Nucleic Acid

1.The concentration and purity of nucleic acid samples, including double-stranded

DNA, single-stranded DNA, RNA, and other nucleic acids.

2.

Microarray samples can simultaneously detect the concentration of nucleic acid and

fluorescent dye, and directly give the concentration value

Protein

A280 measured protein sample concentration, including BSA, IgG, Lysozyme,

Labels;

Instrument parameters

1. Optical path: 1, 0.1, 0.05mm adjustable
2. Minimum sample volume: 1μl
3. Light source: long life xenon lamp
4. Detector type: 2048 linear CCD array
5. Wavelength range: 230, 260, 280 nm
6. Wavelength accuracy: ±1
7. Wavelength resolution: ≤ 2nm (FWHM@Hg 546nm)

Light absorption rate

accuracy: 0.002Abs

8. Light absorption accuracy: 1% (0.76 absorbance at 350nm)
9. Absorbance value range: 0.02-100 (equivalent to 10mm optical path)
10. Concentration detection range: 2~15000ng/μl (DS-DNA) 0.1~100mg/ml

(BSA)

11. Detection period: less than 5 seconds
12. Working voltage: DC 12V/4A
13. Maximum power: 25W
14. Sample base: stainless steel
15. Dimensions: 30cm*20cm*18cm

16.

Weight: 3.5kg

Summary of Contents for DW-K2800

Page 1: ...ation and operation Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd Greeland Max Mall PuDong Shanghai China Tel 0086 21 509660...

Page 2: ...r operations mentioned in this manual Shanghai Drawell Scientific Instrument Co Ltd is not responsible for accidents caused by potentially bad operations not mentioned in this manual If you follow the...

Page 3: ...e performance of electronic components Working environment requirements 1 The instrument should be placed in a dry room with a relative humidity of no more than 85 2 The instrument should be placed on...

Page 4: ...asy to grasp Principle After the spectrophotometer is in the blank the instrument automatically records the spectral result of the blank reference solution and saves it as the light intensity referenc...

Page 5: ...trument parameters 1 Optical path 1 0 1 0 05mm adjustable 2 Minimum sample volume 1 l 3 Light source long life xenon lamp 4 Detector type 2048 linear CCD array 5 Wavelength range 230 260 280 nm 6 Wave...

Page 6: ...platform The instrument has a built in computer and does not require an external computer When the sample is detected the sample can be dropped on the susceptor to form a liquid column and the ultra m...

Page 7: ...system switch for 3 seconds then release the instrument system starts and it will directly enter the test program interface after booting You can use the USB interface to connect an external mouse ke...

Page 8: ...rent samples in order to make the sample form a liquid column better the recommended sample detection amount is as follows the user can adjust according to the actual situation Nucleic acid solution 1...

Page 9: ...an view the absorbance and ratio at 230 260 and 280 nm 4 Seting You can view the type of nucleic acid or protein sample to be detected The nucleic acid has three kinds of dsDNA ssDNA and RNA The prote...

Page 10: ...ta of the next list in the form of an Excel spreadsheet 10 Move Up Move the data strip up 11 Name Name the sample to be measured 12 Key Click to display the soft keyboard 13 Delete Click to delete all...

Page 11: ...nology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd Greeland Max Mall PuDong Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn...

Reviews: