DV-8000 USER’S MANUAL

8

In case the point-to–point measurement shows a wavelength different from the peak

wavelength of the didymium filter (maximum permissible error is ±2nm), remove the
wavelength knob, loosen the three positioning screws on the wavelength dial, adjust
the dial pointer to the characteristic absorption peak wavelength value, and tighten the
screws

3. Troubleshooting

Troubles

Causes

Remedies

1. No functioning after
power on.

1.Powersupply not connected
2.Power fuse melted
3.Loose

contact

of

power

switch.

1. Check power line.
2. Replace fuse.
3. Replace power switch.

2.Unstable display

1.Insufficient warm-up time
2. Excessive vibration, strong air
current near light source or
strong external light.
3. Unstable voltage.
4. Poor grounding.

1. Ensure 20 mins warm-up
time.
2.Improve

working

conditions
3. Use a Voltage Stabilizer.
4.Check ground connection.

3.Unable to set zero

1.black block in the light path
2.Amplifier failure.

1. Remove the block.
2. Repair amplifier.

4.Unable to set 100%T

1.Tungsten-halogen lamp not lit.
2. Inaccurate light path.
3. Amplifier failure.

1. Check or repair lamp
power circuit.
2. Readjust light path.
3.

Repair

or

replace

amplifier.

5.Concentration
incorrectly calculated

1. Display board out of order.

1. Repair or replace display
board.

Summary of Contents for DV-8000 Series

Page 1: ...before installation and operation Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel 0086 21 5...

Page 2: ...ng Principle 4 Unpacking Instructions 4 Specifications 5 Installation 5 Introduction 6 Description of keys and operating method 7 Operating Procedures 11 Maintenance 11 Calibration and trouble shootin...

Page 3: ...es Some of the chemicals used in spectrophotometer are corrosive and or inflammable and samples may be radioactive toxic or potentially infective Care should be taken to follow the normal laboratory p...

Page 4: ...rects the beam onto the grating The grating disperses the light beam to produce the spectrum a portion of which is focused on the exit slit of the mono chromator by a collimating mirror From here the...

Page 5: ...instrument 4 Use the appropriate power cord and plug into a grounded outlet 5 Turn on your DV 8000 model spectrophotometer Allow it to warm up for at least 20 minutes before taking any readings Model...

Page 6: ...to select the Absorbance Transmittance Concentration and Factor mode PRINT key has two functions a To be used in Parallel Port and date transmission one way transmission from the instrument to the Pri...

Page 7: ...T key till it reaches 1500 Press the MODE key again to go to C mode If the A value measures 0 125 then the C value reads 187 5 Operation Procedures 1 Plug the instrument into a grounded outlet 2 Turn...

Page 8: ...ack of the instrument BE SURE NOT TO PULL PANEL WIRING LOOSE 6 Unplug and remove the lamp from ceramic base the white connector Insert the new lamp pushing it in as far as it will go CAUTION a Do not...

Page 9: ...wer line 2 Replace fuse 3 Replace power switch 2 Unstable display 1 Insufficient warm up time 2 Excessive vibration strong air current near light source or strong external light 3 Unstable voltage 4 P...

Page 10: ...echnology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn Email...

Reviews: