
32
The presented figure shows a typical graph showing vag-
inal mucus resistance fluctuations and the optimal mo-
ment for mating.
Often in practice there is a variation between animals
compared to the presented curve and the reading of
the peak value.
For some bitches the peak value is at 400 units, for others
at 600 - 750 units or more than 1,000 units. It should be
remembered, however, that the “profile graph” is the most
important and not the unit values of the readings ob-
tained.
We recommend the following interpretation of readings:
– If the reading is between 100-200 units, there is no need
for daily measurements.
– When readings rise above 200 units, take daily measure-
ments. Sometimes readings at this level can last for sev-
eral days and then show dynamic growth.
– When readings begin to rise, it is recommended that you
perform the tests more frequently (2, 3 or even 4 times
a day) to determine the exact moment of ovulation. This
is especially important for bitches that have ovulation
early and accept males for a short period of time.
As shown in the graph, the most important issue is to cap-
ture the day of maximum resistance and the subsequent
decline. The first and, at the latest, the second day of the
decrease in resistance is the ideal time for mating.
Be aware of the fact that a typical phenomenon of a “false
peak” occurs before the actual peak. It is easy to identify
because it occurs at much lower resistance levels around
day 3 to 5.
SECTION 9 | RESULT INTERPRETATION
Summary of Contents for DOD
Page 3: ...2...
Page 4: ...3 EN INTRODUCTION...
Page 6: ...5 EN EQUIPMENT SECTION 1...
Page 8: ...7 EN DESIGN OF THE DEVICE SECTION 2...
Page 10: ...9 EN SECTION 2 DESIGN OF THE DEVICE APPROXIMATE DIMENSIONS 7 5 cm 8 0 cm 33 0 cm 12 5 cm...
Page 11: ...10 SECTION 2 DESIGN OF THE DEVICE...
Page 12: ...11 EN KEYBOARD FUNCTIONS SECTION 3...
Page 14: ...13 EN STARTING THE DEVICE SECTION 4...
Page 17: ...16 SECTION 4 STARTING THE DEVICE...
Page 18: ...17 EN NOTES TO THE MEASUREMENTS SECTION 5...
Page 21: ...20 SECTION 5 NOTES TO THE MEASUREMENTS...
Page 22: ...21 EN ANIMAL EXAMINATION SECTION 6...
Page 25: ...24 SECTION 6 ANIMAL EXAMINATION...
Page 26: ...25 EN PROBE INSERTION METHOD SECTION 7...
Page 29: ...28 SECTION 7 PROBE INSERTION METHOD...
Page 30: ...29 EN DISINFECTION SECTION 8...
Page 32: ...31 EN RESULT INTERPRETATION SECTION 9...
Page 34: ...33 EN Optimalmomentformating False peak Units Days SECTION 9 RESULT INTERPRETATION...
Page 35: ...34 200 400 600 800 1000 SECTION 9 RESULT INTERPRETATION...
Page 36: ...35 EN SECTION 9 RESULT INTERPRETATION...
Page 37: ...36 SECTION 9 RESULT INTERPRETATION...
Page 38: ...37 EN MAIN MENU SECTION10...
Page 47: ...46 SECTION 10 MAIN MENU...
Page 48: ...47 EN BATTERY REPLACEMENT SECTION11...
Page 50: ...49 EN FINAL REMARKS SECTION12...
Page 52: ...51 EN TECHNICAL DATA SECTION13...