Dräger SAM 3100 Instructions For Use Manual Download Page 8

8

Dräger SAM 3100/3200

Maintenance

6

Maintenance

Dräger SAM 3100/3200 aspirator system:

The aspirator system is designed so as to minimise the need
for servicing.

Directions for maintenance and servicing can be found in
the Instructions for Use for the components used.

Depending on the dust load in the sample gas, the filter
inserts should be replaced when visibly soiled. Spare parts:
see the Instructions for Use for the filter.

In the case of systems with gas sampling pump, it is
recommended that the diaphragm and valve plates of the
pump are replaced at least once a year.

Spare parts: refer to the Instructions for Use for the gas
sampling pump.

The ejectors for delivering sample gas have no mechanical
parts subject to wear, and are therefore maintenance-free.

Gas transmitter

Maintenance and servicing of the gas transmitter should be
carried out according to the enclosed Instructions for Use.

The calibration interval and calibration instructions can be
found in the Instructions for Use for the gas transmitter and
for the DrägerSensor.

For aspirator systems with gas sampling pump, the pump
should be switched off during test gas feeding (calibrating).
For models with ejector, the air feed should likewise
be interrupted.

7

Wiring diagrams

CAUTION

Electrical wiring is only to be laid and connected by an
expert, paying attention to the relevant regulations and
laws concerning electrical equipment in potentially
explosive atmospheres as well as the approval
conditions.

Consult the Instructions for Use for installation!

!

Summary of Contents for SAM 3100

Page 1: ...Dr ger SAM 3100 3200 Sampling system Instructions for Use i...

Page 2: ......

Page 3: ...ystem with ejector unit 5 4 4 System with gas sampling pump 5 4 5 Circuit breaker option 5 4 6 Motor circuit protector option 6 5 Installation 6 5 1 Mechanical fastening 6 5 2 Installing gas lines 6 5...

Page 4: ...ulty or incomplete parts is forbidden The appropriate regulations must be observed at all times when carrying out repairs on these devices or components 1 2 Definitions of alert icons The following al...

Page 5: ...ulating valve on the flow meter The top of the ball is the read mark for the scale Notice on handling see the Instructions for Use for the flow meter and for the gas sampling pump 4 5 Circuit breaker...

Page 6: ...E depending on the chemical resistance to the target gas When using other hose materials or diameters a transition to 6 x 1 mm PTFE hose must be created Cut off hose squarely Assembly instructions for...

Page 7: ...the flow meter 5 3 4 Wiring diagram The electrical connection of the installed intrinsically safe equipment to the and terminals is carried out in accordance with the following illustration and table...

Page 8: ...or delivering sample gas have no mechanical parts subject to wear and are therefore maintenance free Gas transmitter Maintenance and servicing of the gas transmitter should be carried out according to...

Page 9: ...on SE Ex PR DD Power supply 24 V 20 0 15 A 4 to 20 mA 0 mA Controller Unit max cable length 50 m 100 m 150 m core cross section 0 5 mm2 1 0 mm2 1 5 mm2 1 2 3 1 2 3 PEX 3000 all types PEX 3000 types XT...

Page 10: ...tegory 1 classification must be cut out of the rating plate hanger at the location provided Once the equipment has been operated in accordance with this installation it may no longer be installed in m...

Page 11: ...o 20 mA current source source operation Wiring diagram for operation as a 4 to 20 mA current sink sink operation 2123944_en eps RS 1 2 3 4 5 6 power supply red black brown white green yellow central d...

Page 12: ...e base plate Additional equipment Gas sampling pump not for Ex areas PM21772 87 N87 TTE ISW KNF Gas sampling pump for Ex areas PM21773 87 N87 TTE Ex ISW KNF Protective housing 500 x 500 x 300 mm plast...

Page 13: ......

Page 14: ......

Page 15: ......

Page 16: ...o KGaA Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 www draeger com 90 23 944 GA 4675 790 de Dr ger Safety AG Co KGaA Edition 02 September 2013 Edition 01 October 2006 Subj...