Quick Start Guide Dräger P
X
4
5
00
569
5103
1 of 3
General Hints
Every use of a Dräger P
X45
00 requires the exact knowledge and compliance of the instruction
manual 5695
102,
the corresponding norms and standards, the applicable legal provisions, and the
regula-tions and rules for health and safety at work.
Provisional sealing plugs (rubber plugs) of pressurised pipings, that are used over an operating
pressure of 0.5 bar, represent an increased safety risk.
The instrument is only destinated for the applications, as described in the instruction manual. In order to in-
sure the proper function and measurement accuracy, an annual maintenance and re-adjustment has to be
carried out by an authorized service partner.
Before each measurement make sure that the instrument and the accessories are in a perfect condi-
tion.
Charge the Dräger P
X
4
5
00 only with an USB power supply with 5V DC / 1
,5
A.
Instruction manual and PC measurement data
You can download the instruction manual from our website www.draeger-msi.de under the menu Online-
Services
→
Download
→
P
X
4
5
00.
To download the PC measurement data management go to www.draeger-msi.de -> Online-Services ->
Download -> P
X
4
5
00 -> PC200P. The required USB drivers are installed automatically.
Since 2005 across the EU the Waste Electrical and Electronic Equipment Directive
(WEEE 2002/96/EC) have been in force that governs that for private households
garbage collecting and recycling points are arranged. Because the Dräger P
X
4
5
00 isn't registered for
the use in private households, it isn't allowed to dispose it that way. The instruments can be returned to your
na-tional Dräger Safety Organisation for disposal. If you have any questions about the disposal, contact
Dräger MSI GmbH, please.
Switch ON
Press
the
On-/Off Button
for 1 second till it gets illuminated.
Switch OFF
Choose in the main menu 'OFF', or hold the
pess
On-/Off Button
for at least 5 seconds from any menu op-
tion.
Functions
Via the main menu individual tests and measurements can be called up.
Pressure measurements
1.
Fine pressure up to 150 mbar (air and gas)
2.
Pressure up to 3.5 bar (option external sensor; air and gas)
3.
High pressure up to 25 bar (option external sensor; air, gas and water)
Follow the instructions of the Dräger P
X
4
5
00 when testing gas pipes!
The test pressure for performing tightness tests can be pressurized by every standard bike pump or com-
pressor equipped with hose and Schrader valve.
General tightness test
Via this function it is possible to perform tightness tests with freely selectable nominal pressure, stabilisation
time and measuring time.
Select test pressure: 20 mbar – 25,000 mbar
Select stabilisation time: 5 minutes – 6 hours
Select measuring time: 5 minutes – 6 hours
Gas pipe tests