Quick Reference Guide Primus
Infinity Empowered
SW 4.5n
25
Monitoring
Monitoring
Selecting the standard screen
The standard screen is automatically displayed when-
ever a ventilation mode is selected (
Man.Spont.
,
Vol. Mode
or
Press. Mode
).
This screen can always be selected during operation:
z
By pressing the
key (B),
or
z
By pressing the
key (A) repeatedly.
Display (example):
The most important parameters are grouped together
on the right (C) and left sides of the screen.
The three curves are displayed in the middle (D)
Monitoring mode
Monitoring
can be activated in
Standby
, for instance
for exclusive measurement of the SpO
2
value. Fresh-
gas is not delivered.
To start the monitoring mode:
1
Press the soft key
monitor.mode
(A)
or
2
Press the
key (B).
To return to
Standby
:
3
Press the
key (B) again and confirm via the ro-
tary knob.