Dräger Polytron 8100
23
5.8.5
Function key
This function activates the Quick Menu.
The selected information about status or instrument settings
can be retrieved (e.g. warning messages, fault messages,
installed modules, calibration data and time/concentration
graph.)
1. Select
Settings > Instrument > Function key
and
confirm.
2. Select a function and confirm.
Repeating the confirmation will deactivate the selected
function.
In normal operation, the selected function can be activated by
briefly tapping [UP] for 1 to 2 seconds.
If no function key is selected, the Quick Menu is deactivated.
Selection options:
If Bump-test is selected:
The maintenance symbol [ ] is shown on the display and the
maintenance signal is transmitted.
After 15 minutes or tapping [OK] again will end the bump
test and the instrument returns to normal operation.
5.8.6
Device initialization
This function resets the instrument to factory default settings
(see Section 7 on Page 36).
Alarm settings
Passwords
Language
Function key
HART interface
Data-logger
Analog interface
Relay configuration
Bump test
1. Select
Settings > Instrument > Device init.
and confirm.
2. Select
Confirm
and confirm with [OK].
5.8.7
Bump test
A known concentration of gas can be applied to verify sensor
response without setting off the alarms while in the bump test
mode. The maintenance symbol [ ] is displayed. The analog
output is set to the maintenance signal. Pressing [OK] takes
the user our of bump test and back to measurement.
5.8.8
Software dongle
This function deactivates a SW dongle to safely remove the
dongle or if the dongle is faulty.
A dongle can only be reactivated by cycling the power to the
instrument.
1. Select
Settings > Instrument > SW dongle >
Sensortest dongle
or
Diagnostic dongle
confirm.
2. Select
Deactivate func.
and deactivate
Sensortest dongle
or
Diagnostic dongle
with [OK].
5.8.9
Display settings
Changing the contrast
1. Select
Settings > Instrument > Display > Display
contrast
and confirm.
2. Change the contrast and confirm.
Changing the display mode
This function turns the display off during normal operation.
1. Select
Settings > Instrument > Display > Display mode
and confirm.
2. Select
Standard
or
non display
and confirm.
If "non display" has been selected, the display shows the start
screen and the applicable symbols.
If the relay option is installed and an alarm is triggered:
regardless of the selected display mode, the display will show
the current gas concentration and the red LED will blink.
Display test
The function tests the display and the LEDs.
1. Select
Settings > Instrument > Display > Displaytest
and confirm.
2. Select
Enable
or
Disable
and confirm.
When activated, the display is shown inverted and all the LEDs
are lit.
Graph
The measurements of the past 15
minutes are displayed as a time/
concentration graph.
Faults
Fault messages are displayed in plain
text see Section 5.2.1 on Page 18.
Warnings
Warning messages are displayed in
plain text
Bump-test
The bump-test allows applying gas to
the sensor without generating an
alarm. The maintenance signal is
transmitted.
Sensor vitality
(only with
diagnostic dongle)
Shows the remaining sensor vitality
(only available for certain sensors).
Summary of Contents for Polytron 8100
Page 2: ......
Page 3: ...1 5 4 3 2 5 1 4 3 10 7 8 9 2 6 3 4 5 2 1 3 9 5 2 1 14 13 3 4 5 6 7 8 14 2 11 12 15 A B C D E...
Page 223: ...Dr ger Polytron 8100 223 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Page 250: ...250 Polytron 8100 Additional Approvals A 3 Control Drawings...
Page 252: ...252 Polytron 8100 Additional Approvals A 4 Declaration of conformity...
Page 253: ......