Description
Power supply: U
i
= 30 V, I
i
= 0.3 A, P
i
= 700 mW, C
i
= 5 nF, L
i
= 50
μ
H
Year of manufacture (via serial number)
1)
Dräger Safety, 23560 Lübeck, Germany
Type P3U
Safety-related characteristic values for the supply and signal circuit (outer terminals
of the docking station):
Ui = 30 V, I
i
= 0.3 A, P
i
= 700 mW, C
i
= 5 nF, L
i
= 50
μ
H
Type P3FB
FISCO bay unit
Power supply: U
i
= 24 V, I
i
= 0.38 A, P
i
= 5.32 W, C
i
= 5 nF, L
i
= 10
μ
H
3.5.2 IECEx
Types P3U & P3FB
IECEx BVS 04.0003X
Year of manufacture (via serial number)*
Dräger Safety, 23560 Lübeck, Germany
Type P3U
Power supply: U
i
= 30 V, I
i
= 0.3 A, P
i
= 700 mW, C
i
= 5 nF, L
i
= 50
μ
H
Ex ia I Ma
-40 °C
≤
Ta
≤
+65 °C
(-40 °F
≤
Ta
≤
+149 °F)
II 3G
Ex ic IIC T4 Gc -40 °C
≤
Ta
≤
+65 °C
(-40 °F
≤
Ta
≤
+149 °F)
Ex ic IIC T6 Gc -40 °C
≤
Ta
≤
+40 °C
(-40 °F
≤
Ta
≤
+104 °F)
1) Structure of the serial numbers: The third letter of the serial number indicates the year of man-
ufacture: M = 2019, N = 2020, P = 2021, R = 2022, S = 2023, T = 2024, U = 2025,
W = 2026, X = 2027, Y = 2028, Z = 2029 etc. (letters G, I, O, Q are omitted)
Example: Serial number ARMB-0001: The third letter is M, meaning that the device was man-
ufactured in 2019.
0158
Ex ia IIC T4 Ga
-40 °C
≤
Ta
≤
+65 °C
(-40 °F
≤
Ta
≤
+149 °F)
Ex ia IIC T6 Ga
-40 °C
≤
Ta
≤
+40 °C
(-40 °F
≤
Ta
≤
+104 °F)
Ex ia I Ma
-40 °C
≤
Ta
≤
+65 °C
(-40 °F
≤
Ta
≤
+149 °F)
Ex ic IIC T4 Gc
-40 °C
≤
Ta
≤
+65 °C
(-40 °F
≤
Ta
≤
+149 °F)
Ex ic IIC T6 Gc
-40 °C
≤
Ta
≤
+40 °C
(-40 °F
≤
Ta
≤
+104 °F)