Use
4.4.2 Passwords
For setting the passwords, see section 9.3.
4.4.3
Opening the menu
●
To open the
Information
directly: In measuring mode, press and hold down
for approx. 1 second.
●
To open the
Information
and
Calibration
menus:
a. In measuring mode, press and hold down for approx. 3 seconds.
b. Select
Enter password
.
c. Enter the calibration password.
●
To open all menus:
a. In measuring mode, press and hold down for approx. 3 seconds.
b. Select
Enter password
.
c. Enter the settings password.
4.4.4
Reading out information
In the menu
Information
choose the appropriate menu item:
Confirms inputs.
Selects menus and functions.
Key
Function
OK
Calibration password
Access to
–
Information
menu
–
Calibration
menu
Factory setting:
_ _ _ 1
Settings password
Access to
–
Information
menu
–
Calibration
menu
–
Settings
menu
Factory setting:
_ _ _ 2
M
M
M
Instrument
Show Notice
Shows existing warnings. For troubleshooting with respect
to the warning number displayed
(e.g. #58) see section 7.2.
If multiple warnings are present, this is indicated (e.g. 1/3 =
page 1 of 3.).
Show Fault
Shows existing errors. For troubleshooting with respect to
the error number displayed (e.g. #100) see section 7.1.
If multiple errors are present, this is indicated (e.g. 1/3 =
page 1 of 3.).
Show Fault
Codes
Shows present error and warning codes in a table. If all
numerical groups are displayed with "00", no error is pres-
ent.