Assembly instructions
|
Docking Station for Ex e version of Polytron® 5000/8000
en
|
Safety information
Docking Station for Ex e version of Polytron® 5000/8000
1
Safety information
Any use of the Docking Station for the e-version of Polytron
1)
5xx0/8xx0
requires full understanding and strict observance of the instructions for use of
the respective Dräger Polytron instrument (e. g. Dräger Polytron 5000/8000
instructions for use).
2
Conventions in this document
2.1
Meaning of the warning notes
The following warning signs are used in this document to indicate and highlight
areas of the associated text that require particular attention by the user. A
definition of the meaning of each sign is as follows:
3
Functional description
The Docking Station is required for the electrical installation in any area with an
"increased safety" explosion protection rating.
● Instrument for increased safety (Ex e)
The instrument is extended by an increased safety terminal box (docking
station) that provides up to four 20 mm openings, which can be used for field
wiring or wiring of a remote sensor. The permissible cable diameter range is
7 to 12 mm.
The Docking Station can be pre-mounted, wired and sealed with the supplied
cover. Once the site is ready for commissioning, the instrument is then hooked
up to the Docking Station and taken into operation; avoiding that the instrument
is damaged during the construction phase.
The connection between the Docking Station and the main instrument is
realized via a ‘feedthrough’. Depending on the instrument selected, there are 3
types of feed-through.
– feedthrough 3 wires for power
– feedthrough 9 wires for power and relay
– feedthrough 14 wires for power, interface connections or relay and remote
sensor (not for Polytron
®
8100 EC)
4
Intended use
The Docking Station is intended for installation either in hazardous areas of
zone 1, 2 or 21, 22 according to device categories 2G, 3G or 2D, 3D.
The Docking Station is suitable for wall, ceiling and duct mounting.
Duct mounting requires additional mounting kits.
5
Assembly
The gas detector can be used to measure gases in the ambient air or in a
pipe/duct.
1. Attach the Docking Station to one of the following options.
1) Polytron® is a registered trademark of Dräger.
Alert icon
Signal word
Consequences in case of nonob
-
servance
WARNING
Indicates a potentially hazardous situa
-
tion. If not avoided, it could result in
death or serious injury.
CAUTION
Indicates a potentially hazardous situa
-
tion. If not avoided, it could result in
physical injury. It may also be used to
warn against unsafe practices.
NOTICE
Indicates a potentially hazardous situa
-
tion. If not avoided, it could result in
damage to the product or environment.
Option
Mounting on a flat surface
Use a drilling template (part number
45 44 299) and attach it using 4
screws (M6 x 10).
Mounting to a bar or beam
Mount kit (part number 45 44 198)