background image

PAS

®

 Lite Series

Self contained breathing apparatus

Instructions for Use

i

PAS

®

 is a registered trademark of Dräger

3357781 (A3-D-P) Page 1 of 2

1

For your safety

1.1

General safety statements

Before using this product, carefully read the Instructions for Use.

Strictly follow the Instructions for Use. The user must fully understand

and strictly observe the instructions. Use the product only for the

purposes specified in the Intended Use section of this document.

Do not dispose of the Instructions for Use. Ensure that they are

retained and appropriately used by the product user.

Only fully trained and competent users are permitted to use this

product.

Comply with all local and national rules and regulations associated

with this product.

Only trained and competent personnel are permitted to inspect, repair

and service the product. Dräger recommends a Dräger service

contract for all maintenance activities and that all repairs are carried

out by Dräger.

Properly trained service personnel must inspect and service this

product as detailed in the Maintenance section of this document.

Use only genuine Dräger spare parts and accessories, or the proper

functioning of the product may be impaired.

Do not use a faulty or incomplete product, and do not modify the

product.

Notify Dräger in the event of any component fault or failure.

The air supply shall meet the requirements for breathing air according to

EN12021

.

1.2

Definitions of alert icons

Alert icons are used in this document to provide and highlight text that

requires a greater awareness by the user. A definition of the meaning of

each icon is as follows:

WARNING

Indicates a potentially hazardous situation which, if not avoided,

could result in death or serious injury.

CAUTION

Indicates a potentially hazardous situation which, if not avoided,

could result in physical injury or damage to the product or

environment. It may also be used to alert against unsafe practices.

NOTICE

Indicates additional information on how to use the product.

2

Description

2.1

Product overview

The Dräger PAS

®

 Lite Series is a self contained breathing apparatus

(SCBA) that uses an open-circuit breathing system. The series is

compatible with a wide range of compressed-air cylinders, face masks and

lung demand valves (e.g. FPS 7000 and Panorama Nova face masks, PSS

Series lung demand valves, and steel or composite cylinders).

2.2

Feature description

The carrying system uses a lightweight polymer-composite space frame

(Fig 1, Item 3). The shoulder straps and waist belt are fully adjustable and

use webbing harness straps with rubberized shoulder straps and a plastic

waist belt buckle.

The apparatus uses the Dräger high-performance pressure reducer (8)

that reduces cylinder pressure and supplies breathing air to the lung

demand valve (1) through a medium-pressure hose (7). Medium-pressure

air is supplied either directly or through a coupling (9) that allows removal

and fitting of the lung demand valve when required. A gauge hose (5)

supplies high-pressure air from the cylinder to the gauge (4). The reducer

also incorporates a whistle (6) that sounds to warn the wearer that there is

low cylinder pressure.

The air hoses and the pressure reducer are integrated into the space frame

to prevent snagging and enhance component protection. The hose clips

(2) hold the air hoses in position on the shoulder straps.

Air cylinders, lung demand valves and face masks

The Dräger PAS Lite Series is compatible with a single steel or composite

material cylinder of 4 to 9 litre capacity, and 200 or 300 bar pressure. Full

descriptions and user instructions are contained in separate instructions

supplied with the cylinder, mask or lung demand valve.

2.3

Intended use

The PAS Lite Series breathing apparatus, when fitted with a cylinder, lung

demand valve and face mask, is intended for use by emergency services

and in industrial applications where a high level of respiratory protection is

required. The assembled breathing apparatus provides the wearer with

respiratory protection for working in contaminated or oxygen-deficient

conditions, and is suitable for fire fighting.

The cylinder, face mask and other accessories used with this product must

be certified Dräger components, assembled in an approved configuration.

Contact Dräger for further information.

Use in potentially explosive atmospheres

Electronic sub-assemblies are ATEX certified.

2.4

Approvals

The European standards, guidelines, and directives according to which

this product is approved are specified in the declaration of conformity (see

declaration of conformity or www.draeger.com/product-certificates).

2.5

Explanation of marking and symbols

Refer to the relevant authority for explanation of approval body symbols

and marking on the equipment. Examples of other marking on component

parts of the breathing apparatus are:

BRAC-1359 

Dräger serial number

08/09

Month and year of manufacture

3356812 or R21034  –

Dräger part number

SF

Standard force coupling

LF

Low force coupling

!

!

i

i

3

Use

WARNING

Only trained and competent personnel may prepare and use

breathing apparatus. Ensure that any accessories, ancillary

equipment and other protective clothing items do not interfere with

the apparatus and do not create a safety hazard.

The effective working duration of the apparatus is dependent on

the initial air supply available and the breathing rate of the wearer.

Fill air cylinders to their full rated pressure prior to use, and do not

commence any operation using a cylinder that is less than 90 per

cent full (or greater when national regulations dictate).

CAUTION

Do not apply excessive force or use tools to open or close a
cylinder valve, and do not drop or throw down the breathing
apparatus.

3.1

Preparation for use

NOTICE

The pressure gauge may be fitted with a thin flexible protective
covering. Remove this covering before first use.

1. Carry out a visual inspection of the breathing apparatus (see

Section 3.5.1).

2. Fit the air cylinder (see Section 3.5.2).

3. For breathing apparatus with a medium-pressure coupling for the lung

demand valve, disconnect and then reconnect the male coupling. To
connect, press the male coupling into the female coupling until an
audible click is heard. If there is any difficulty disconnecting or
connecting, see the troubleshooting information in Section 4.

4. Press the reset button (Fig 1, Item 11) when using a positive-pressure

lung demand valve (see the Instructions for Use supplied with the lung

demand valve).

5. Carry out a full functional test of the apparatus (see Section 3.5.3).

6. Connect the lung demand valve to the face mask and check the

security of attachment by gently attempting to pull the coupling apart.

3.2

Putting on the breathing apparatus

1. Fully loosen the shoulder straps and waist belt and put on the

breathing apparatus.

2. Check that the shoulder straps are not twisted and take the weight of

the system on the shoulders by pulling the shoulder straps. Do not fully

tighten at this stage.

3. Close the waist belt buckle and pull the ends of the waist belt until it fits

securely and comfortably on the waist (Fig 2). Tuck the belt ends

behind the waist belt.

4. Pull the shoulder straps until the breathing apparatus is secure and

comfortable. Do not over tighten. Tuck the belt ends behind the waist

belt.

!

!

i

i

5. Fully loosen the head straps of the face mask and place the neck strap

over the back of the neck.

6. Press the reset button when using a positive-pressure lung demand

valve.

7. Open the cylinder valve slowly, but fully, to pressurize the system.

8. Put on the face mask and check for tight fit (see the Instructions for Use

supplied with the mask).

3.3

During use

WARNING

Fully open the cylinder valve and ensure that it remains open

during use.

Users should be in a safe area before the whistle warning

commences. Evacuate to a safe area immediately if the warning

commences during an operation.

Using the supplementary air supply (Fig 1, Item 10) will use air
from the cylinder and rapidly reduce the working duration of the
apparatus.

Regularly check the remaining cylinder pressure on the gauge.

If supplementary air is required, briefly press the rubber cover (10) at

the front of the lung demand valve to deliver extra air into the face

mask.

3.4

After use

WARNING

Do not remove the breathing apparatus until in a safe breathing
environment.

1. Loosen the face mask straps. As the seal to the face is broken, press

the reset button when using a positive-pressure lung demand valve.

Remove the face mask.

2. Close the cylinder valve.

3. Press the rubber cover at the front of the lung demand valve to vent

the system fully. Press the reset button when using a positive-pressure

lung demand valve.

4. Release the waist belt buckle.

5. Lift the shoulder strap buckles to loosen the straps.

6. Remove the breathing apparatus and face mask.

7. Carry out the after use tasks in the maintenance table (see Section 5).

8. Remove the air cylinder (see Section 3.5.2) if required.

9. Pass the breathing apparatus to the service department with details of

any faults or damage that occurred during use.

3.5

Common user tasks

3.5.1

Visual inspection

A visual inspection must check the full breathing apparatus including all

component parts and accessories. Check that the equipment is clean and

undamaged, paying particular attention to pneumatic components, hoses

and connectors. Typical signs of damage that may affect the operation of the

breathing apparatus include impact, abrasion, cutting, corrosion and

discolouration. Report damage to service personnel and do not use the

apparatus until faults are rectified.

3.5.2

Air cylinder fitting and removing

WARNING

High-pressure air release may cause injury to the user or other

personnel near the breathing apparatus. Close the cylinder valve and

fully vent the system before attempting to disconnect an air cylinder.

Impact damage to the cylinder valve or reducer connector may

prevent valve connection or cause an air leak. Handle the air

cylinder and breathing apparatus with care.

NOTICE

For other cylinder connector types, refer to the Instructions for Use
supplied for the connector.

Fitting a cylinder with a threaded connector

1. Check the threads of the cylinder valve port and the pressure reducer.

Ensure that the O-ring seal (Fig 3, Item 1) and the sintered filter (2) in

the reducer are clean and undamaged.

2. Lay the apparatus horizontal, with the reducer uppermost, and fully

extend the cylinder strap.

3. Insert the cylinder through the loop of the strap, and align the valve with

the reducer.

4. Lift the cylinder and space frame into the vertical position (supported

on the end of the cylinder opposite the valve).

5. Tighten the hand wheel of the reducer, using only the thumb and index

finger, until a definite metal-to-metal contact is felt. Do not use tools or

over tighten.

6. Place the unit back into the horizontal position.

7. Take up the slack in the cylinder strap (Fig 4).

8. Pull the strap over the cylinder to operate the cam lock (Fig 5).

9. Secure the strap end using the cylinder strap retainer (Fig 6). Release

the cam lock to adjust the position of the cylinder strap retainer if

necessary.

Removing a cylinder with a threaded connector

1. Close the cylinder valve and fully vent the system.

2. Lay the apparatus horizontal, with the cylinder uppermost.

3. Lift the cylinder strap retainer.

4. Lift the strap against the cam lock to release the buckle tension, and

then loosen the strap.

5. Disconnect the cylinder valve from the pressure reducer.

6. Lift the cylinder away from the reducer and remove the cylinder.

3.5.3

Functional testing

WARNING

Failure of the breathing apparatus to meet any of the standards or
parameters described in the functional tests indicates a system fault.
Report the fault to trained service personnel or contact Dräger. Do
not use the breathing apparatus until the fault condition is rectified

.

Assemble the breathing apparatus as described in the preparation for use

(see Section 3.1) before commencing any functional testing.

!

!

!

i

i

!

WRAH-1653

COMP

RESSED A

IR

bar

0

50 
100

150200

250

300

'

'

Dräg

er

Dräg

er

Dräge

r

Dräge

r

1

2

1

3

4

5 6

7
8

9

11

10

3115

5

'

100

150

50

200

250

300

0

bar

CO

MPRESSED AIR

'

2

3110

3108

6

3

1

2

3111

3104

4

3109

Reviews: