DRAFT October 2 2006
4.
Connect standard tubing (not provided) to the airway pressure port (F) (see figure 2 on
page 4) and to the pressure port on the manual bagging apparatus (G) (see figure 1 on
page 3).
NOTE:
If manual bagging apparatus are used without a pressure port, then use an adapter with
a side pressure port part number MU10902.
NOTE:
Resuscitation Modules without an airway pressure port should use disposable breath-
ing circuit part number MU10841.
5.
Connect the manual bagging apparatus (E) to the apparatus at the patient, such as
mask, tent, nasal cannula, endotracheal tube.
Figure 2. Resuscitation Module (without AutoBreath™ Infant Resuscitator): Controls
WARNING:
Ensure the patient breathing circuit connections are secure and free of
obstructions. Failure to do so could result in patient injury.
WARNING:
During infant resuscitation, monitor the airway pressure, and provide
appropriate pressure relief. Failure to do so could result in patient injury.
Suction
Suction
Off
On
0
Patient Supply
Patient Outlet
Flow Rate (LPM)
Airway Pressure Relief
0
Patient Outlet 2
Flow Rate 2 (LPM)
Airway Pressure
5
10
0
15
5
10
0
15
C
F