Dräger X-am 5100
21
Use
3.3
Performing the bump test
1. Prepare a test gas source (e.g., cylinder, permeation oven);
the volume flow must be 0.5 L/min and the gas concentration must
be higher than the alarm threshold concentration to be tested.
2. Fit the calibration adapter
(68 06 291) to the sensor cap.
3. Connect the test gas source
to the calibration adapter.
4. Vent the test gas into a fume
cupboard or into the open air
(with a hose connected to the
second connector of the
calibration adapter).
5. Switch on the instrument.
6. Open the valve on the test
gas source to let test gas flow
over the sensor.
7. Recommendation: Wait until
the instrument displays the test gas concentration with acceptable
tolerance.
However, wait at least until alarm threshold A1 or A2 has been
exceeded.
If the alarm thresholds are exceeded, the instrument displays the
gas concentration alternately with
» A1 «
or
» A2 «
depending on
the test gas concentration.
8. Close the valve on the test gas source.
If the concentration has now fallen below the A1 alarm threshold:
Acknowledge the alarm.
If gas concentration display is of acceptable tolerance:
Calibrate the instrument, see “Calibrating the instrument” on
page 25.
WARNING
Never inhale the test gas. Danger to health!
Observe the hazard warnings in the relevant Safety Data
Sheets.
NOTICE
Sensor-specific features affecting the bump test are
described in the respective sensor data sheets.
!
i
i
x-am 5100
0
00533280.eps
Summary of Contents for MQG 0020
Page 2: ......
Page 28: ...28 Dr ger X am 5100 Technical Data...
Page 69: ...Dr ger X am 5100 69 Dados t cnicos...
Page 158: ...158 Dr ger X am 5100 1 6 170 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Page 163: ...Dr ger X am 5100 163 3 1 5 1 2 Bump Test 21 O2...
Page 166: ...166 Dr ger X am 5100 166 3 6 3 6 1 A1 A1 A1 1 2 3 6 2 A2 A1 i i OK...
Page 170: ...170 Dr ger X am 5100 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr gerSensor 6 2...
Page 220: ...220 Dr ger X am 5100 1 6 232 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Page 225: ...Dr ger X am 5100 225 3 1 5 1 2 Bump Test 21 O2...
Page 228: ...228 Dr ger X am 5100 9 A1 3 5 228 3 6 3 6 1 A1 A1 A1 A1 A2 3 6 2 A2 A1 i i OK...
Page 232: ...232 Dr ger X am 5100 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr ger 6 2 7 Dr ger Dr ger...
Page 259: ...Dr ger X am 5100 259 1 6 271 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Page 264: ...264 Dr ger X am 5100 3 1 4 3 2 1 3 1 5 1 2 Bump Test Bump Test OK 21 Vol O2...
Page 270: ...270 Dr ger X am 5100 5 5 6 001 7 8 9 10 11 12 OK 13 5 14 OK OK OK OK OK OK...
Page 271: ...Dr ger X am 5100 271 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr gerSensor 6 2...
Page 286: ...286 Dr ger X am 5100 Notes on Approval A 2 Declaration of Conformity...
Page 287: ......