14-14
D
ELTA
/D
ELTA
XL/K
APPA
VF8
14 I
NVASIVE
B
LOOD
P
RESSURE
Labeling Pressure Channels
The pressure label determines how a signal is analyzed and reported to the monitor.
For detailed information about the various types of pressure labels, see page 14-2.
When you assign a new label to a pressure channel, the monitor clears the parameters
and conditions set for the previous label (including alarms and waveform scales) and
replaces them with settings for the new label. Trends are stored according to the
assigned label.
Zero
,
Cal Factor
, and
Cal Date & Time
settings are associated with
the pressure channel and are kept even during a label change.
The
Pressure Labels
screen can display up to twelve IBP sources in a 3
X
4 matrix. The
monitor assigns an automatic pressure label (
P
[
1
-
3
][
a
-
d
]) to each box.
Signal sources are displayed in rows [1-3], with Row 1 representing data received
from the HemoMed pod or Y-cable, and Rows 2 and 3 representing data from the
Hemo2, Hemo4, MPod – Quad Hemo or PiCCO pod, either of which may be
Manometer Cal
Allows you to enter the manometer or
simulator reading and start calibration
(see page 14-5).
10 – 300
Pressure Overlap
Allows you to view up to four IBP
parameters on a single baseline.
• ON
• OFF
Common Scale
Sets the waveforms to one scale.
• OFF, 5, 10, 15, 20, 25, 30, 35,
40, 50, 75, 100, 125, 150, 175,
200, 225, 250, and 300
Large Mean
Increases size of all IBP mean values.
• ON
• OFF
Wedge Start
(PA setup menu
only)
Starts a wedge pressure measurement
Same function as the Wedge fixed key
on a hemodynamic pod (see page
NOTE:
Not available on the Picco pod.
Not applicable
Min Scale
(ICP, ICP2, ICP3,
ICP4 setup menus
only)
Sets the ICP waveform scale to +/- 25
mmHg (+/- 3.3 kPa).
NOTE:
When the
Min Scale
setting is
set to
ON
, the
Scale
selections are
unavailable.
• ON
• OFF
Pressure Labels
Displays the Pressure Label screen
(see page 14-14).
Not applicable
<IBP Parameter>
Alarms
Displays alarms for parameter and
associated variable on Alarm Limits
table.
Not applicable
IBP Setup Menus
Menu Item
Description
Available Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...