24 | © 2015
•
Increase O
2
Function
−
Avea: FiO
2
increases by 10%
Increase FiO
2
Button
−
Dräger: FiO
2
increases by 1.3 times the set FiO
2
Example 1: Set FiO
2
is 25%. O
2
Suction is activated, FiO
2
will increase to 33%
Example 2: Set FiO
2
is 60%. O
2
Suction is activated, FiO
2
will increase to 78%.
O
2
Suction Button
o
Push O
2
Suction on screen
o
Confirm by pushing control knob
180 seconds of oxygen increase
Differences in Increase O
2
or
O
2
Suction