Short Operating Instructions Dräger FG4200 

 

5695050 

 

 

Edition 04 – January 2019 

 

 

en

 

General safety instructions

 

 

Any use of the FG4200 requires full understanding and compliance with the operating 
instructions 5695051, the relevant standards, as well as the relevant statutory regulations.

 

 

The device is intended only for the uses described in this manual. Any improper use of the device may lead to 
electric shock or destruction of the measuring instrument!

 

 

Before using the measuring device for the first time, the integrity of all components must be checked, e.g.:

 

 

Device exhibits no visible damage 
No condensed water in the gas preconditioner 
The gas preconditioner filter is clean 
Gas hoses without defects 
Visual inspection of the probe

 

 

Always fully charge the FG4200 via the USB interface with a 5 V DC / 1 A USB power supply 
adapter only.

 

Incomplete charging affects the charging capacity of the battery in the long term.

 

For longer periods of 

inactivity, we recommend a monthly charge for at least 8 hours. When charging, no measurements should be 
performed. 

 

Operating Instructions and managing PC measuring data

 

The operating instructions and the measurement data management can be found on our website 

http://www.draeger-msi.de/home/?L=1

 under the menu item 

Online-Services

 → 

Download 

→ 

FG4200

.  

The necessary USB drivers are installed automatically by downloading the PC data management. 
 

Since 2005, EU-wide regulations apply to the disposal of electrical and electronic equipment. In 
essence, these regulations set out that collection and recycling options must be available for private 
households. Because the FG4200 is not registered for use in private households, it must not be 

disposed of through such channels. The devices can be returned to your national retailer or to your national 
Dräger Safety Organization for disposal. Please contact Dräger MSI GmbH, if you have any questions 
regarding disposal. 
 

Maintenance 

To ensure proper function and accuracy, calibration and adjustment should be performed once a year. Device 
maintenance must only be carried out by trained service personal. 

 

Switching on

 

Press the display slightly for 1 second, then press 

Next

 

 

Switching off

 

Select and activate the 

OFF

 menu item in the main menu or press the display for 5 seconds whilst in any 

menu point.

 

 

Make sure that the gas outlet on the side is unobstructed and not closed or blocked! 

 

Functional areas

 

The individual tests and measurements are accessed via the main menu.

 

 

Flue gas measurement

 

To perform a complete flue gas measurement, we recommend a measurement time of at least 2 minutes. In 
addition to the flue gas measurement, it is also possible to calculate averages and measure the draft and 
enter the combustion system date. 

 

Switch on the device and wait for system check. Then connect the flue gas probe to the unit (see connection 
diagram). 

 

Ambient air CO measurement 

Measurement of the CO concentration in ambient air. The zero point adjustment has to be done with fresh air. 