![Dräger D-Vapor Instructions For Use Manual Download Page 71](http://html.mh-extra.com/html/drager/d-vapor/d-vapor_instructions-for-use-manual_2526743071.webp)
Instructions for use D-Vapor/D-Vapor 3000
71
Operation
4
Turn the anesthetic agent bottle against the
noticeable spring resistance until a higher
noticeable resistance is reached.
The bottle valve is now open. Hold the bottle firmly
in this position. Desflurane flows from the tank into
the anesthetic agent bottle.
5
Drain until the tank is empty or the bottle is full.
6
Slowly turn the anesthetic agent bottle (B)
counterclockwise.
The spring in the integrated closure (C) helps
the bottle valve close. The draining procedure is
completed.
7
Position the vaporizer upright again.
8
Continue to slowly turn the anesthetic agent
bottle and withdraw it from the filling port.
To remove the residual desflurane left in the
vaporizer, see "Blowing off the vaporizer"
on page 71.
If necessary, repeat the procedure with an empty
anesthetic agent bottle.
Ending the draining
1
Place the screw cap (A) in the filling port and
screw it all the way in.
2
Close the anesthetic agent bottle with the cap.
3
Mark the bottle:
"Used anesthetic agent."
Recommendation: Do not reuse.
Blowing off the vaporizer
If the anesthetic agent remaining in the tank also
has to be removed after complete draining:
1
Connect the vaporizer to the mains power
supply and put it into operation.
2
Set the control dial to 18 Vol% and flush with
10 L/min air until the
Delivery Low
alarm is
triggered.
3
Allow gas to flow into the scavenging line.
4
Press the
0
button and set the control dial to the
T
position.
Plug-in adapters: engage locking lever in
control dial.
5
Disconnect the vaporizer from the mains power
supply.
082
WARNING
Spraying of anesthetic agent
If the bottle continues to be turned after the
bottle valve has been opened, anesthetic
agent may spurt out with considerable force.
Do not continue to turn.
0
Vo
l.%
Summary of Contents for D-Vapor
Page 14: ...14 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 15: ...Instructions for use D Vapor D Vapor 3000 15 Use Use Intended use 16 Environment of use 16 ...
Page 26: ...26 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 76: ...76 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 88: ...88 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 114: ...114 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 116: ...116 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 119: ...Instructions for use D Vapor D Vapor 3000 119 This page has been left blank intentionally ...