16.
Remove XRAY tray….using recessed handle. Pull out XRAY tray as far as it will go. Then tilt the drawer upwards and pull out completely:
Page 1: ...erilzing a done after each patient change b see Operator s Manual for list of approved cleaning disinfectants 2 Check air filter Tilt unit for easy access 3 View of filter area Part number MX 17 015 4...
Page 2: ...the Cleaning Mode 6 ALLOW CALEO TO COOL DOWN after self cleaning one hour 7 Remove canopy Grasp the handles on the sides of the canopy with both hands Lifting the canopy horizontally 8 Wipe surfaces w...
Page 3: ...Remove all grommets 12 Clean the grommets with a disinfecting solution 13 Remove the mattress from inside the incubator 14 Wipe mattress down with approved cleaner and check for holes or punctures in...
Page 4: ...16 Remove XRAY tray using recessed handle Pull out XRAY tray as far as it will go Then tilt the drawer upwards and pull out completely...
Page 5: ...17 Wipe down all surfaces with approved disinfectant 18 The remaining tray is also removed 19 Remove that remaining piece by pressing the two catches inwards and lifting up at the same time...
Page 6: ...in plain view remove PLEASE note maybe HOT 22 Remove any soiling and wipe down with approved cleaning material 23 After drying replace impeller 24 Turn over trough part removed to expose fan impeller...
Page 7: ...HE CONTROL PANEL DO NOT DISINFECT THE CONTROL PANEL BY IMMERSION OR SPRAYING ________________________________________________________________________ Summary of the parts to be cleaned _______________...