6.5.3  Fitting the power pack

 WARNING

Exposed battery terminals might cause a risk of explosion or fire through 
sparking.

Only fit a telemetry version (Fig. 5, Item 1) to a breathing apparatus 
that is fitted with a PSS

®

 Merlin

®

 Modem.

Only fit a non-telemetry version if the terminal cover (Fig. 5, Item 3) is 
secure and undamaged.

1. Inspect the power pack and pressure module, and ensure that the 

battery terminals and sealing rim are clean and undamaged. Contact 
service personnel or Dräger if there is any damage.

2. Insert the power pack into the backplate recess (Fig. 7).

Fig. 7

Inserting the power pack

3. Position thumbs on top of the 2 screws and push down firmly to lock 

the power pack (Fig. 8).

Fig. 8

Pressing the power pack down

4. While pushing down, confirm that the 2 sliding latches move to their 

locked position viewed through the 2 keyholes as shown (Fig. 9).
– When the power pack connects, a tone sounds and the start-up 

sequence commences (see section 4.4.3).

Fig. 9

Checking the latches

5. Telemetry version (Fig. 5, Item 1) only: tighten (nip up) the screw 

using a suitable coin until it bottoms in the recess and a noticeable 
increase in turning force is felt. Do not over tighten or damage will 
occur.

6.5.4  Replacing 1.5 V batteries

 WARNING

Explosion, fire, or chemical hazard.

Do not remove or install the batteries in an explosive or flammable 
atmosphere.

Do not expose the batteries to heat sources, do not attempt to 
recharge any non-rechargeable battery, and do not short out the 
battery terminals.

Use only the recommended battery type, replace batteries as a 
matching set, and do not mix new and used batteries.

NOTICE

Environmental hazard. Dispose of used batteries in accordance with 
national or local regulations. More information can be obtained from local 
waste disposal organizations.

Use only the following approved battery types.

– Duracell

®

 LR6 (1.5 V)

– Duracell

®

 Plus LR6 (1.5 V)

1. Remove the power pack (see section 6.5.2).
2. Remove the 8 screws (Fig. 10) using a 2.5 mm hexagon key.

Fig. 10

Replacing 1.5 V batteries

3. Remove the battery cover.
4. Remove the batteries.
5. Install a new set of batteries observing the polarity marked inside the 

pack.

6. Inspect the sealing ring inside the battery cover. Contact Dräger or 

service personnel if sealing ring replacement is necessary.

7. Refit the battery cover and tighten the screws. Do not over tighten 

(Dräger recommend tightening to 0.4 Nm (0.3 lbf ft)).

6.5.5  Charging 6.5 V rechargeable power packs

Rechargeable power packs can be charged using the Dräger 4-Way 
Charger for both versions, or the Dräger In-Cab Charger for the telemetry 
version only. For charging procedures, see the 4-Way Charger or PSS

®

 

Merlin

®

 instructions for use.

7 Storage

– Store the product in accordance with the associated breathing 

apparatus instructions for use.

– If the system will not be used for a long period:

– Remove the power pack (see section 6.5.2).
– If the power pack has replaceable 1.5 V batteries, remove the 

batteries from the power pack (see section 6.5.4).

8 Disposal

8.1 General

Dispose of the product in accordance with applicable rules and 
regulations in the country of use.

8.2  Waste electrical and electronic equipment 

(WEEE) directive

9 Technical 

data

10  Manufacturer and document 

information

 Manufacturer

Dräger Safety UK Limited

Ullswater Close
Blyth, NE24 4RG
United Kingdom
Tel: +44 1670 352 891
Fax: +44 1670 356 266

www.draeger.com

3356513

_

en

© Dräger Safety UK Limited
Edition: 14 – 2019-08 (Edition: 1 – 2007-12)
Subject to alteration

41347

41348

41349

41356

This product must not be disposed of as household waste. This is 
indicated by the adjacent symbol.
You can return this product to Dräger free of charge. For informa-
tion please contact the national marketing organizations or Dräger.

EOST alarm

Commences

60 bar to 50 bar

Stops

10 bar to 8 bar

Power supply

Replaceable battery type

5 x 1.5 V battery

Rechargeable type

6.5 V NiMH

DSU

Pre-alarm activation

21 to 25 seconds

Full alarm activation

Approximately 8 seconds

Nominal frequency and power level

125 kHz

66 dB

μ

A/m at 10 m

40 kHz

42 dB

μ

A/m at 10 m