background image

24

Instructions for use

  |  

Dräger BG ProAir

  

Maintenance 

5.2.6.10 Checking the pre-flush unit

(only for apparatuses with on-demand oxygen dosage)

The pre-flush unit only needs to be checked for 

apparatuses with on-demand oxygen dosage. This chapter 

does not apply for apparatuses with constant oxygen dosage.

1. Set the selector switch on the test unit to 

.

2. Slowly fully open the cylinder valve.

Oxygen must be heard to flow into the circuit. The pre-

flush unit must automatically shut-down after max. 

10 seconds.

5.2.6.11 Performing the high-pressure leak test (incl. 

constant dosage device)

For apparatuses with constant oxygen dosage, the 

constant dosage device is also automatically checked during 

the high-pressure leak test. 

1. Set the selector switch on the test unit to 

.

2. Slowly open the cylinder valve by at least two turns.

The Connect ECU automatically switches to On mode.

An acoustic signal sounds.

The battery status is shown on the display.

The Connect ECU queries whether to carry out the high-

pressure leak test (

). If the query is confirmed, the 

 

symbol and the current pressure are displayed.

A high-pressure leak test can only be carried out if the 

cylinder pressure is at least 150 bar and the Connect ECU 

is in On or Standby mode.

3. Close the cylinder valve.

For apparatuses with on-demand oxygen dosage, use the 

Connect ECU to confirm that the cylinder valve has been 

closed.

The test starts automatically for apparatuses with constant 

oxygen dosage.

The test is carried out.

The result of the test is shown after approx. 15 seconds. 

The following results may occur:
– The closed-circuit breathing apparatus is OK. The 

 

symbol is shown on the Connect ECU display.

– An error has occurred. The 

 or 

 symbol is shown 

on the Connect ECU display.

Repeat the test or have the apparatus serviced.

If, for national or regulatory reasons, the constant 

dosage device needs to be checked with a test unit, 

see chapter 5.2.7.1.

5.2.6.12 Checking the bypass

1. Set the selector switch on the test unit to 

.

2. Slowly open the cylinder valve by at least two turns.

The Connect ECU automatically turns on and runs a self-

test.

3. Briefly press the bypass button.

Oxygen must be heard to flow into the circuit (flow noise).

5.2.6.13 Checking the minimum valve

1. Set the selector switch on the test unit to 

.

2. Swing the locking lever into the spring bridge and make 

sure that it is latched in position.

3. Observe the display on the test unit and wait until the 

pressure stops dropping.

4. Read the pressure on the display.

The value must be b0.5 mbar and +3.0 mbar.

5.2.6.14 Checking low-pressure alarm 1

1. Set the selector switch on the test unit to 

.

2. Close the cylinder valve.
3. In apparatuses with on-demand oxygen dosage, gently 

press the bypass button so that the pressure in the 

breathing circuit drops.

The pressure automatically drops for apparatuses with 

constant oxygen dosage.

4. Observe Connect ECU: 

At approx. 55 bar, the colour on the display switches to red 

and low-pressure alarm 1 should be triggered.

5. Remove the connecting piece from the test unit connector.

The apparatus is completely purged.

6. Fit the protective cap on the connecting piece.
7. Turn off the Connect ECU. To do so, simultaneously press 

the right and left button until 3-2-1 is shown on the display.

The Connect ECU turns off.

8. Turn off the test unit.

5.2.7  Maintenance work on components

5.2.7.1  Checking the constant dosage device 

(only for apparatuses with constant oxygen dosage)

Work equipment

– RZ 7000 (R62500)
– The following parts from the BG ProAir Test Set 

(3722266):

(1) Quick connect coupling (3722211)

(2) Test hose (E02120)

(3) Adaptor RZ 7000 (R62599)

The constant dosage device only needs to be checked for 

apparatuses with constant oxygen dosage. This chapter does 

not apply for apparatuses with on-demand oxygen dosage. 

The constant dosage device is automatically checked during 

the high-pressure leak test. If necessary, it can also be 

checked separately at the Dräger RZ 7000.

1. Creating the test setup:

a. Switch on the test unit.
b. Connect the test hose (2) to the test unit using the 

adaptor RZ 7000 (3).

c. Connect the test hose with the quick connect coupling 

(1) to the medium-pressure hose for the oxygen supply 

at the CO

2

 absorber (see overview graphic at start of 

instructions, item 7).

2. Swivel the spring bridge upwards in order to disable the 

function of the minimum valve.

Summary of Contents for BG ProAir

Page 1: ...Instructions for use Dr ger BG ProAir More languages available for download at draeger com ifu Dr ger BG ProAir...

Page 2: ......

Page 3: ...200 0 02 bar h elapsed time 1 4 2 3 27 27 11 16 9 10 8 29 28 24 25 26 14 15 12 13 19 20 23 22 21 5 6 7 17 18 31 30 8 7 6 9 1 2 3 4 5...

Page 4: ...digitalt exemplar p andra spr k i databasen f r teknisk doku mentation www draeger com ifu SV 9300712 ET Selle kasutusjuhendi saab teistes keeltes elektroonili sel kujul alla laadida tehnilise dokume...

Page 5: ...use 12 4 3 1 Freezing the ice packs for apparatuses with ice cooler 12 4 3 2 Inserting ice packs for apparatuses with ice cooler 12 4 3 3 Donning the closed circuit breathing apparatus 12 4 3 4 Activ...

Page 6: ...to potential hazards The meanings of the warning notices are defined as follows 2 2 Trademarks The following website lists the countries in which the trademarks from Dr ger are registered www draeger...

Page 7: ...g cooler options are available Ice cooler The ice cooler contains ice packs that can be refrozen after use Regeneration cooler The regeneration cooler PCM phase chage material cooler contains a latent...

Page 8: ...following information is shown when the compressed oxygen cylinder is opened Other information and queries are shown if the Connect ECU is turned on differently pushbutton insertion of batteries remo...

Page 9: ...mmersion of the apparatus in water e g when navigating a waterhole in a mine will not impair the function of the closed circuit breathing apparatus If the closed circuit breathing apparatus has been c...

Page 10: ...ion on approvals see supplement 9300233 3 6 Type identifying markings and explanation of symbols 3 6 1 Type identifying marking The type identifying marking of the BG ProAir Class 4P is located on the...

Page 11: ...ored for 6 months The apparatus must have successfully passed the leak test and be sealed with the protective cap The air tight shrink wrapped or installed CO2 absorber was not transported on a vehicl...

Page 12: ...ing aid to prevent the ice packs from deforming Make sure that no ice is deposited on the ice packs If necessary place ice packs in a bag seal the bag and freeze the ice packs whilst they remain in th...

Page 13: ...g the automatic distress signal function Remove the function key from the Connect ECU This activates the motion sensor The symbol is displayed 4 4 During use 4 4 1 General information The chemical pro...

Page 14: ...ecess when tightening the bayonet ring Rotate the bayonet ring clockwise onto the connection to fasten the breathing hose The breathing hose is correctly fitted when a clicking sound is heard while ti...

Page 15: ...k the catches on the fasteners under the tabs see Figure B 1 Push the fasteners closed see Figure B 2 Connecting piece The direction of flow is marked with an arrow on the connecting piece The directi...

Page 16: ...g circuit may contain base which may leak when disassembling the closed circuit breathing apparatus and lead to chemical burns Avoid skin contact Wear protective gloves and goggles In case of eye cont...

Page 17: ...th release levers 1 Swivel the spring bridge 90 and remove it from the closed circuit breathing apparatus 2 Release the locking mechanisms between the CO2 absorber and breathing bag as well as between...

Page 18: ...4 Prepare the CO2 absorber Dispose of the disposable absorber For the refill absorber remove the soda lime and the two non woven mats and discard them The sieve can be reused after disinfection 5 2 3...

Page 19: ...rmation 5 2 5 Assembling the closed circuit breathing apparatus 5 2 5 1 Overview 1 Install the CO2 absorber see Installing the CO2 absorber page 19 2 Install the breathing air cooler see Installing th...

Page 20: ...nect the red medium pressure hose to the breathing air cooler The quick connect coupling must engage When connecting make sure that a click is heard If necessary pull on the hose and check whether the...

Page 21: ...th the locking mechanisms to the CO2 absorber and breathing air cooler Every locking mechanism has two lock buttons with engagement hooks Ensure the locking mechanisms on the CO2 absorber and breathin...

Page 22: ...en the Connect ECU on the shoulder strap 5 2 6 Testing the closed circuit breathing apparatus 5 2 6 1 Notes for testing The following test sequence applies for both apparatuses with constant oxygen do...

Page 23: ...he selector switch on the test unit to 2 Tightly pinch the exhalation hose with your hand Use the face spanner from the BG ProAir Test Set if necessary 3 Pump once The test unit must display at least...

Page 24: ...ically turns on and runs a self test 3 Briefly press the bypass button Oxygen must be heard to flow into the circuit flow noise 5 2 6 13 Checking the minimum valve 1 Set the selector switch on the tes...

Page 25: ...b 400 when handling the CO2 absorber WARNING Reduced period of use possible If the expiration date of the soda lime is exceeded the intended period of use for the apparatus may not be reached Note the...

Page 26: ...s are shown on the display 7 Confirm the battery type 8 Turn off the Connect ECU 6 Transport Condensate forms in the breathing circuit during use After use transport the closed circuit breathing appar...

Page 27: ...5 L min Bypass 80 L min Minimum valve 80 L min Temperature during continu ous use1 1 according to EU type examination 6 C to 30 C Air pressure 900 hPa to 1200 hPa Relative humidity 0 100 Storage temp...

Page 28: ...ylinder 2 200 O2 CGA USA R360395 BG ProAir 2 200 O2 Configurator R360560 Accessories Dr ger FPS 7000 RP R56326 Dr ger FPS COM 5000 R62700 klar pilot anti fogging spray R56542 BG ProAir disposable abso...

Page 29: ...tightness test with positive pressure Lower pressure to 7 0 5 mbar Press OK button Pressure difference 1 mbar min Engage lever in spring bridge Pump to activate Opens between 4 and 8 mbar Relief valve...

Page 30: ...e pressure Lower pressure to 7 0 5 mbar Press OK button Pressure difference 1 mbar min Relief valve Engage lever in spring bridge Pump to activate Opens between 4 and 8 mbar Bypass Open cylinder Brief...

Page 31: ...Annex Instructions for use Dr ger BG ProAir 31...

Page 32: ......

Page 33: ......

Page 34: ...9300708H Manufacturer Dr ger Safety AG Co KGaA Revalstra e 1 D 23560 L beck Germany 49 451 8 82 0 9300708 1114 800 en Dr ger Safety AG Co KGaA Edition 01 2021 10 Subject to alterations www draeger com...

Reviews: