Dräger Babylog V Series Disassembly/Assembly Download Page 5

D

-5755-2018

Dräger Babylog

®

 V-Series 

Dräger Neonatal Flow Sensor

Disinfection

Disinfection

At the beginning of the contact time, swirl the components back and forth at least 3 times. Make sure the disinfectant 

reaches all surfaces and interiors. At the end of the contact time, swirl the components back and forth again 

at least 3 times.

After the contact time, rinse the components in the water bath (at least drinking water quality) until no disinfectant residue is visible.

Place the components in the disinfectant. Observe the specified contact time.

Shake out residual water carefully. 

Allow the components to dry com-

pletely.

Manufacturer:

Drägerwerk AG & Co. KGaA
Moislinger Allee 53–55
23558 Lübeck, Germany

This guideline does not replace or amend the instructions for use. Please always refer to the 

reprocessing instruction in the latest instruction for use of the components. The instructions 

for use and the hospital guidelines and policies always have to be observed.