Disassembly
/Assembly
Neonatal Expiratory Valve
Open the flap by lifting the lower
edge upwards.
Close the flap.
Turn the locking ring to the left
end position.
Turn the locking ring to the right
until it clicks into place.
Remove and empty the water
trap container.
Remove the muffler from the
expiratory valve.
Turn the locking ring as far as
possible to the left and push the
expiratory valve into the fitting.
Remove the expiratory valve from
the fitting.
Remove the diaphragm.
Do not disassemble further.
91
03
108
|
19.11-1
|
A
W
|
LE
|
Subject
to
modifications
|
©
2019
Dr
äg
er
w
er
k
A
G
&
C
o.
K
G
aA
Attach the diaphragm onto the edge
of the expiratory valve housing.
Make sure that it fitts properly.
Attach the muffler to the
expiratory valve.
Connect the water trap container.
Dräger Babylog
®
V-Series
Neonatal Expiratory Valve
Disassembly/Assembly
Disassembly
/Assembly
Neonatal Expiratory Valve
The muffler is only used with the Babylog
®
V-Series.
6
6
5
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.